User Guide
MES 1003 WMS
Doc no. 100837-MAN – Rev. E1 – Rev. date 2022-10
© Danfoss IXA A/S | Page 37
Illustration 41: Immerse the sensor into pH 7 buffer solution
The pH sensor should be immersed in pH 7 buffer solution after the sensor has been cleaned in distilled water.
Illustration 42: pH sensor setup for calibration of pH
The container with pH buffer solution can be placed on top of the measurement chamber and the pH sensor can be
inserted in the buffer solution. The setup can be seen in above illustration.
Press OK to start the calibration.
Illustration 43: Calibrating the pH sensor
Wait for the calibration to finish
pH Calibration
6-**
Calibrating pH 7
Immerse the sensor into pH
7 buffer solution
Press OK when done
pH Calibration
6-**
Calibrating…