Piranha4 8K Color Camera User's Manual
33
Teledyne DALSA
03-032-20156-00
Verify Temperature and Voltage
To determine the voltage and temperature at the camera, use the
Refresh Voltage and Refresh
Temperature
features found in the
Camera
Information
set.
The temperature returned is the internal temperature in degrees Celsius. For proper operation, this value
should not exceed 75 °C. If the camera exceeds the designated temperature it will stop imaging and the
LED will turn red. Once you have diagnosed and remedied the issue use the
reset camera
function.
The voltage displayed is the camera‘s input voltage.
Note:
The voltage measurement feature of the camera provides only approximate results (typically within
10% and dependent on the voltage drop in the cable). The measurement should not be used to set the
applied voltage to the camera, but only used as a test to isolate gross problems with the supply voltage.
Saving and Restoring Camera Settings
The parameters used to select, load and save user sets are grouped together under the Camera
Information set of features. There are 8 user sets available and one factory set.
Camera Information
Parameter
Choices
User Set Default Selector
Select the camera parameters to load when the camera is reset or
powered up as the Factory set, or as User Set 1 to 8.
Selecting the set from the list automatically saves it as the default
set.
User Set Selector
Select the Factory or User set to Save or Load.
-Factory Set
-User Set 1 to 8.
User Set Load
Load the set specified by User Set Selector to the camera and make
it the active / current set.
User Set Save
Save the current set as selected user set.
Description of the Camera Settings
The camera operates in one of three settings:
1.
Current session.
2.
User setting.
3.
Factory setting (read-only).
4.
Default setting.
The current settings can be saved (thereby becoming the user setting) using the User Set Save parameter.
A previously saved user setting (User Set 1 to 8) or the factory settings can be restored using the User Set
Selector and User Set Load parameters.