DAIHAN Scientific, Rotary Vacuum Evaporator
25/31
-
Temperature Offset Function
is used to set up an offset value and thereby add or deduce a
certain value when the temperature of the thermometer used in a user’s laboratory needs to be
equal to the temperature displayed on Digital bath controller.
-
An available offset value ranges from -10.0
℃
to +10.0
℃
. It can be adjusted in the unit of 0.1
℃
.
-
The current temperature after an offset value is set up means the temperature including the
offset value.
-
If Jog-dial knob is pushed long after a temperature offset value is set up, the set value flickers
and is saved.
-
If you go to Set Temp Offset Mode while Heating operation is executed, Heating stops. After
setting up an offset value, execute Heating operation again in Set Temperature Mode.
-
If Mode button is pushed in Set Temp Offset Mode, it is possible to change to Set Temperature
Mode.
③
Locking Mode
-
If Mode button is pushed once in Set Temperature Mode, it is possible to change to Locking
mode.
-
In Locking Mode, it is impossible to turn or push Jog-Dial Knob. The current operation status is
displayed on LCD.
-
If Mode button is pushed once, it is possible to go to Set Timer Mode.
※
After operation, it is recommended to use Locking Mode for test safety.
Fig 3.3.5. Set Temp Offset Mode Screen
Current Temp.
Offset Temp.
PV
26.8
- 10.0
SV
℃
℃
Heater
Operation status
Timer
Remainin
Current Temp.
PV
26.8
: 10
℃
Fig 3.3.6. Bath Locking Mode Screen