4.5.2
UV Flow Cell ÄKTApilot 1/2/5 mm and Industrial Flow Cell
8 mm and 1 inch
Installing the flow cell
Follow the steps below to install UV Flow Cell ÄKTApilot 1/2/5 mm and Industrial Flow
Cell 8 mm and 1 inch.
Step
Action
1
Unpack the flow cell. Do not remove the black protective caps on the optical
fiber transmitter and receiver connector.
2
Choose the optical path length of the flow cell. The optical path length can
be changed with shim plates.
3
Attach the flow cell to an appropriate flow cell holder.
4
Connect the inlet and outlet tubing to the flow cell.
Connecting the optical fibers
Follow the steps below to install the optical fibers.
WARNING
High intensity UV light. This product uses high intensity ultra-
violet light. Do not disconnect the optical fibers while the lamp is
on.
NOTICE
Do not touch the tips of the optical fiber with anything other than
lens paper.
NOTICE
To avoid damaging the optical fibers, press only on the cell body,
never on the optical fibers.
4 Installation
4.5 Installing the flow cells
4.5.2 UV Flow Cell ÄKTApilot 1/2/5 mm and Industrial Flow Cell 8 mm and 1 inch
Monitor UVis-920 Operating Instructions 29055049 AF
37