Parameter
Specification
Chemical resistance
The wetted parts are resistant to
organic solvents and salt buffers
commonly used in chromatography of
biomolecules, except 100% ethyl
acetate, 100% hexane, and 100% tetra-
hydrofuran (THF) or 15% THF in acetoni-
trile
Tubing connections
UNF 10-32 fingertight connectors for
tubing with 1/16" outer diameter
UV Flow Cell ÄKTApilot 1/2/5 mm
Parameter
Specification
Recommended maximum flow rate
800 mL/min
Maximum pressure
2 MPa (20 bar, 290 psi)
Backpressure
Max. 0.05 bar at 400 mL/min with water
at 25°C
Liquid temperature range
4°C to 40°C
Optical path length
1, 2 and 5 mm
Cell volume
300 μL
Wetted materials
PEEK (polyetheretherketone)
Titanium (palladium alloy)
FFKM (perflorelastoner) or EPDM
Quartz (synthetic fused silica)
pH stability range
1-13, 13-14 (< 1 days exposure)
Chemical resistance
The wetted parts are resistant to
organic solvents and salt buffers
commonly used in chromatography of
biomolecules, except 100% ethyl
acetate, 100% hexane, and 100% tetra-
hydrofuran (THF)
Tubing connections
5/16"
8 Reference information
8.1 Specifications
Monitor UVis-920 Operating Instructions 29055049 AF
75