46
TRUSCULPT OPERATOR MANUAL
D1395, REV. K, 02/18
Recommended Separation Distances Between Portable and Mobile RF
Communications Equipment and the truSculpt System
The truSculpt system is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The user of the truSculpt system can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the truSculpt
system as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power (W) of transmitter
Separation distance (m) according to frequency of transmitter
150 kHz to 80 MHz
d = 1.17
√
P
80 MHz to 800 MHz
d = 1.17
√
P
800 MHz to 2.5 GHz
d= 2.33
√
P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
Содержание truSculpt
Страница 1: ...Operator Manual...
Страница 4: ...iv TRUSCULPT OPERATOR MANUAL D1395 REV K 02 18...
Страница 8: ...viii TRUSCULPT OPERATOR MANUAL D1395 REV K 02 18...
Страница 36: ...28 TRUSCULPT OPERATOR MANUAL D1395 REV K 02 18...
Страница 58: ...50 TRUSCULPT OPERATOR MANUAL D1395 REV K 02 18...
Страница 66: ...B 2 TRUSCULPT OPERATOR MANUAL D1395 REV K 02 18...