17
TRUSCULPT OPERATOR MANUAL
D1395, REV. K, 02/18
READY and STANDBY Modes
Except during actual treatment, place the system in STANDBY mode by touching the yellow
(Standby) button on the RF Handpiece Control screen. When the system is in STANDBY mode, the
Standby button illuminates yellow; the Target Temperature and Pulse Duration fields are highlighted in
yellow; and the RF handpiece on/off button is disabled. If the console is not used for 3 minutes, the
system automatically enters STANDBY mode.
To place the system in READY mode, touch the green
(Ready) button. When the system is in
READY mode, the Ready button illuminates green; the Target Temperature and Pulse Duration fields are
highlighted in green; and the RF handpiece on/off button is enabled. When the on/off button is pressed,
RF energy delivery is initiated.
RF Handpiece Control Screen (READY Mode shown)
WARNING
Except during actual treatment, the system must always be in
STANDBY mode. Maintaining the system in STANDBY mode
prevents accidental RF energy exposure.
Standby button
Ready button