Curtiss-Wright | PG Drives Technology
SK81935-01
62 | R-net Omni2 Technical Manual | Chapter 3 – Programming
2.8.4 SWITCH LONG
This sets the period of time the User Switch must be operated, in order to put the Omni2 into Sleep Mode.
The programmable range is 0s to 5.0s in steps of 0.25s.
When using Single Switch Scanner-Type SIDs the Omni2 can be put to sleep via the User Menu or Sequence.
2.8.5 SWITCH MEDIUM
This parameter is intended for 3-Axis Proportional and Switch-Type SIDs only
A direction change occurs at the instant the User Switch is operated, provided the parameter Double Click has been set to 0. If
the User Switch is released and a Forward/Reverse command is entered, then drive will commence in the newly-selected
direction.
If the User Switch is not released and is held for the time period set by the programmable parameter, Switch Medium, then
this is interpreted as a conventional User Switch operation, i.e. the User Menu will be entered or the other powerchair
functions would be sequenced. There will be no direction toggle in this instance.
As with the other two methods, if the User Switch is operated for a period greater than the time set by the programmable
parameter, Switch Long, then the Omni2 will go to sleep.
The programmable range is 0s to 5.0s in steps of 0.25s.
2.8.6 SWITCH DEBOUNCE
This sets the debounce time for the User Switch. Debounce is the period of time the switch is operated continuously before a
new condition is registered.
The programmable range is 50ms to 500ms in steps of 10ms.
2.8.7 DOUBLE CLICK TIME
This sets the period of time during which two User Switch operations must be detected, in order to be interpreted as a
double-click.
The programmable range is 0s to 2.5s in steps of 0.1s.
If set to 0s, then no double-click facility will be supported. This setting should be used if accidental double-clicks are a
problem.
If set to any other value, then this is the period that two User Switch operations must be made, in order to be interpreted as a
double-click.
2.8.8 ACTIVE DOUBLE CLICK
This feature allows the user to set and test the Double Click speed.
This feature is only available in OBP.
Содержание R-NET OMNI2
Страница 1: ...PG DRIVES TECHNOLOGY R NET OMNI2 TECHNICAL MANUAL SK81935 01 ...
Страница 8: ...Curtiss Wright PG Drives Technology SK81935 01 8 R net Omni2 Technical Manual ...
Страница 28: ...Curtiss Wright PG Drives Technology SK81935 01 28 R net Omni2 Technical Manual Chapter 1 Installation ...
Страница 52: ...Curtiss Wright PG Drives Technology SK81935 01 52 R net Omni2 Technical Manual Chapter 2 Operation ...
Страница 74: ...Curtiss Wright PG Drives Technology SK81935 01 74 R net Omni2 Technical Manual Chapter 3 Programming ...
Страница 102: ...Curtiss Wright PG Drives Technology SK81935 01 102 R net Omni2 Technical Manual Chapter 5 Infrared Control ...
Страница 118: ...Curtiss Wright PG Drives Technology SK81935 01 118 R net Omni2 Technical Manual Chapter 6 Bluetooth Set Up Operation ...
Страница 147: ...Curtiss Wright PG Drives Technology R net Omni2 Technical Manual Chapter 9 Specifications 147 SK81935 01 ...
Страница 152: ...Curtiss Wright PG Drives Technology SK81935 01 152 R net Omni2 Technical Manual Chapter 9 Specifications ...