Arrhythmia Detector Performance
ECG Analysis System - ECG Database Test
ECG
Rhythm
Class
Rhythms
Minimum test
sample size
Performance
goal
Test
sample
size
Shock
Decision
No Shock
Decision
Observed
Performance
90% One
Sided Lower
Confidence
Limit
Coarse
VF
200
>90%
sensitivity
219 213 6
97.26%
(213/219)
sensitivity
95%
SH
OC
KABL
E
Fast VT
50
>75%
sensitivity
137 111 26
81.02%
(111/137)
sensitivity
76%
Normal
Sinus
Rhythm
100
minimum
(arbitrary)
> 99%
specificity
100 0 100
100%
(100/100)
specificity
97%
AF,SB,S
VT, heart
block,
idioven-
tricular
PVC’s
30
(arbitrary)
> 95%
specificity
219 1 218
99.54%
(218/219)
specificity
98%
N
ON S
H
OC
KA
BL
E
Asystole 100
> 95%
specificity
132 5 127
96.21%
(127/132)
specificity
93%