2.4 Batteries
The i-PAD NF1200 is equipped with LiMnO2 battery cells. These batteries are nonrechargeable. The status of the
battery charge is monitored by periodic tests (daily, weekly, monthly), power ON test, run-time test, and battery
insertion test. The i-PAD NF1200 indicates through the State LED and the i-Button if the battery pack needs to be
replaced.
If the voltage level of the internal Battery pack falls below the minimum tolerable level, the i-PAD NF1200 is
going to be inoperative.
If this is detected during automatic self-test, the i-PAD NF1200:
a.
Flashes the State LED in red
b.
Emits a burst of beeps every minute.
When the i-PAD NF1200 is turned ON after a low battery condition is detected, the i-PAD NF1200:
a.
Turns the State LED momentarily in blue.
b.
Then turns the State LED in solid red.
c.
Flashes .the i-Button in red
d.
Gives the prompt: “Press the flashing red i-Button”.
e.
After the i-Button is pressed, the i-PAD NF1200 gives the prompt: “Low battery. Replace the battery with a
new one.”
The sequence of indications from a to e above is also given when the low battery level is detected during battery
insertion test.
•
Keep the battery pack out of children’s reach.
•
If the battery pack, leaks and the leaked liquid gets in the eyes, wash them with clean water and
consult a physician immediately.
•
Do not leave the battery pack in direct sunlight or in high temperature areas.
•
Do not have the battery pack in contact with water.
•
Keep the battery pack away from direct sunlight, high temperature, and humidity.
•
Follow local regulations when disposing of the battery pack.
•
Do not subject the battery pack to conditions beyond the safe environmental conditions for the i-PAD
NF1200.
•
Do not charge the battery pack
•
Do not open the case of the battery pack. Do not saw off or break apart the case of the battery pack.
•
Do not let the battery pack come into contact with open flames and other hot objects. Do not dispose
of in fire
•
Do not short-circuit the terminals of the battery pack.
•
Do not subject the battery pack to serious physical impact. Do not hit it with a hammer.
•
In case of leakage or strange smell, keep away from fire to prevent ignition of any leaked electrolyte.
Содержание i-PAD NF1200
Страница 1: ...1 Service Manual CU Medical Systems Inc i PAD NF1200 ...
Страница 8: ...8 Perspective view of the i PAD NF1200 with its cover open ...
Страница 25: ...25 i PAD NF1200 Rescue Protocol ...
Страница 39: ...39 8 You will see the top case after you lift up the main PCB board This is shown in the picture below ...
Страница 41: ...41 6 Assembly Diagram ...
Страница 42: ...42 Assembly Diagram Larger version A3 size ...