18
3.2 Automatic Self-Tests
When the i-PAD NF1200 conducts an automatic self-test (Daily, Weekly, Monthly), it wakes up automatically and
turns the State LED solid blue during the duration of the test. It goes back to standby mode after the test.
3.3 Battery Insertion Test
Insert a functioning battery pack to initiate a battery insertion test.
3.3 Self Test Results
3.4.1 Automatic Self Tests
When no problem is detected during automatic self tests, the i-PAD NF1200 does nothing but increment the test
counter. On the other hand, when a problem is detected, the i-PAD NF1200 does the following:
a.
Upon detecting the problem
i.
Beeper emits bursts of three successive beeps. Bursts are spaced one minute apart.
ii.
State LED flashes in red.
b.
When the i-PAD NF1200 is turned ON while the State LED is flashing in red:
i.
The State LED turns solid blue then solid red.
ii.
The i-Button flashes in red.
iii.
The beeper emits a short beep and the i-PAD NF1200 gives the voice prompt: “Press the flashing red i-
Button”
iv.
After the i-Button is pressed, the beeper emits a short beep and the i-PAD NF1200 gives either of the
following voice prompts: “Low battery level. Replace the battery with a new one.” – if the i-PAD
NF1200 detected a low battery level.
“System failure. Error code is 2”. – if the i-PAD NF1200 detected a system failure. The error code is a
number from 2 to 128.
v.
The i-PAD NF1200 turns off automatically.
vi.
The State LED flashes in red every 5 seconds and the beeper emits three successive beeps every minute
to signify the occurrence of an error.
3.4.2 Battery Insertion Test
If no problem is detected, the i-PAD NF1200 activates the following indicators and voice prompts:
a.
Beeper beeps for 1 second. State LED turns ON in blue for the duration of the self-test.
b.
Shock button flashes in orange. At the same time, the i-PAD NF1200 gives the voice prompt: “Press the
flashing orange button, now”. The voice prompt is looped until the shock button is pressed. Beeper emits a
short beep when the flashing shock button is pressed.
c.
After the shock button is pressed, the i-Button backlight flashes in blue. At the same time, the i-PAD NF1200
gives the voices prompt: “Press the flashing blue i-Button”. The voice prompt is looped until the i-Button is
pressed. Beeper emits a short beep when the flashing i-Button is pressed.
d.
After the i-Button is pressed, the State LED turns OFF then flashes in green three times to signify the end of
the battery insertion test. The State LED then flashes every 5 seconds to signify that it is in standby mode.
Содержание i-PAD NF1200
Страница 1: ...1 Service Manual CU Medical Systems Inc i PAD NF1200 ...
Страница 8: ...8 Perspective view of the i PAD NF1200 with its cover open ...
Страница 25: ...25 i PAD NF1200 Rescue Protocol ...
Страница 39: ...39 8 You will see the top case after you lift up the main PCB board This is shown in the picture below ...
Страница 41: ...41 6 Assembly Diagram ...
Страница 42: ...42 Assembly Diagram Larger version A3 size ...