Failure: Automatic periodic self- tests are not running
Possible Causes
Action
1
Battery pack is not inserted into the i-
PAD NF1200 during storage
Verify that the battery pack is inserted into the i-PAD
NF1200 during storage. Turn the i-PAD NF1200 and verify
that it turns ON.
Insert the battery pack if it is not inserted.
2
Real time clock battery is dead.
The real time clock chip is powered by a separate 3V
Toshiba Coin Battery (CR1220) (BT1 on the main PCB). If
the coin battery voltage falls below 2.5 V, replace it with a
new one.
3
Real time clock chip failure
Replace the main board.
Verify the solution by making the i-PAD NF1200 conduct
automatic self-tests.
If the problem persists, put the original main board back
and contact the manufacturer.
Failure: The i-PAD NF1200 keeps incorrect date and time.
Possible Causes
Action
1
Real time clock chip battery failure.
Replace the dedicated real time clock battery cell.
Verify the solution by doing the following:
1.
Synchronize the i-PAD NF1200 with a clock
using the CU Expert.
2.
Turn the i-PAD NF1200 OFF.
3.
After one hour, determine the time setting of
the i-PAD NF1200 using the CU Expert. Verify
that the time indicated is synchronized with the
clock used in step 1.
2
Real time clock chip failure;
Microprocessor failure
Replace the main PCB. Verify the solution by doing the
verification steps in possible cause no. 1 above.
Содержание i-PAD NF1200
Страница 1: ...1 Service Manual CU Medical Systems Inc i PAD NF1200 ...
Страница 8: ...8 Perspective view of the i PAD NF1200 with its cover open ...
Страница 25: ...25 i PAD NF1200 Rescue Protocol ...
Страница 39: ...39 8 You will see the top case after you lift up the main PCB board This is shown in the picture below ...
Страница 41: ...41 6 Assembly Diagram ...
Страница 42: ...42 Assembly Diagram Larger version A3 size ...