11
1.4.2 Voice and Text Prompts
Prompts
The i-PAD NF1200 guides you throughout a rescue operation using voice prompts and other audio-visual
indicators. The following table lists the voice prompts of the i-PAD NF1200.
Table: Voice prompts and their meanings.
Prompt Meaning/Definition
Attach pads
•
Indicates that you have to attach the defibrillator electrode
pads to the bare chest of the patient and to the i-PAD
NF1200
Do not touch the patient
•
Indicates that you or any bystander must not touch the
patient.
Analyzing heart rhythm
•
Indicates that the i-PAD NF1200 is analyzing the ECG of
the patient. Nobody must touch the patient.
Shock advised
•
Indicates that a shockable ECG rhythm has been detected
and the i-PAD NF1200 is preparing for a shock delivery.
Stand clear
•
Emphasizes the need to stay clear of the patient. Given just
before the prompt to deliver shock is given.
Press the flashing orange button, now.
Deliver shock, now.
•
Indicates that you have to press the shock button in order
for the shock to be delivered. The shock will only be
delivered if you press the shock button.
Shock delivered
•
Indicates that a shock has been delivered to the patient.
Begin CPR, now
•
Indicates the start of CPR.
•
You may touch the patient after this prompt is given.
Push the chest down, fast, two inches
•
Indicates that you have to push the chest down for about
two inches during CPR. The chest has to be pushed down
fast.
Give two breaths. Breath, breath
•
Indicates that you have to give artificial breaths to the
patient. Time the breaths with the breathing guide.
The shock button was not pressed
•
Indicates that you did not press the shock button within 15
seconds when the i-PAD NF1200 prompted you to do so.
No shock advised
•
Indicates that the i-PAD NF1200 detected a nonshockable
rhythm.
Check pulse
•
Indicates that you have to check the pulse of the patient.
If no pulse, begin CPR
•
Indicates that if there is no pulse, you have to begin CPR.
System shutting down
•
Indicates that the ON/OFF button of the i-PAD NF1200
has been pressed to turn it OFF.
Press the flashing orange button, now
•
Given during battery insertion test. Indicates that you have
to press the SHOCK button so that the i-PAD NF1200 can
test its functionality.
Press the flashing blue i-Button
•
Given during battery insertion test. Indicates that you have
to press the i-Button button so that the i-PAD NF1200 can
test its functionality.
Administration mode
•
Indicates that the i-PAD NF1200 is in administration mode
Содержание i-PAD NF1200
Страница 1: ...1 Service Manual CU Medical Systems Inc i PAD NF1200 ...
Страница 8: ...8 Perspective view of the i PAD NF1200 with its cover open ...
Страница 25: ...25 i PAD NF1200 Rescue Protocol ...
Страница 39: ...39 8 You will see the top case after you lift up the main PCB board This is shown in the picture below ...
Страница 41: ...41 6 Assembly Diagram ...
Страница 42: ...42 Assembly Diagram Larger version A3 size ...