CU-ERT Operator’s Manual_ver 1.10
60
APPENDIX A
<TECHNICAL SPECIFICATIONS>
ECG Monitor (Simulated)
• Patient connection:
Defibrillation Pads, ECG Electrodes
• Bandwidth:
Monitoring Mode: 0.3 to 40 Hz (-3 dB)
AED mode (EMS): 1 Hz to 30 Hz
• ECG size:
Auto-scaled (0.3 to 1 mV signals are displayed with 10mm/mV
gain, outside of that range, the peak to peak value is displayed
as 10 mm on the LCD display)
• Heart rate:
Digital readout on LCD display, 30 to 300 bpm (
±
2%)
Defibrillator (Simulated)
• Operating modes:
Semi automatic
• Waveform:
e
-
cube
biphasic (Truncated exponential type)
• Energy:
• AED mode
150J into a 50
Ω
load
• Charging time:
Less than 10 seconds
• Sensitivity & Specificity:
Meets AAMI DF39 guidelines
• Detection Level
> 0.1 mV
• Defibrillation Electrodes:
Multifunction, self-adhesive, adult electrodes (Disposable, Pre-
gelled)
Voice & Text Prompts
• Voice & text prompts guide the user through the rescue protocol
• All the user interfaces are supported in local language for easy use and access
Data Storage & Management
• Internal flash memory:
12 hours of event and ECG recording
• SmartMedia Card (32M):
48 hours of event and ECG recording
or 1 hour recording if voice recording is enabled
• Review of the patient’s ECG, incident details and device information
• Data management software for storage, review, and printing of multiple patient rescue data
using a PC running on Windows OS.
Display
• Screen Type:
High resolution graphic Liquid Crystal Display
• Screen size:
4 inches (10.16 cm) diagonal, 320X240 pixels
• Sweep speed:
25mm/sec, nominal
• Viewing time:
3.2 seconds
Physical
• Size:
305 mm X 250 mm X 95 mm (L X W X H)
• Weight
Approximately 1.4 kg; without the external battery pack
AC Adapter
• Input:
100 to 240V AC 50/60Hz 1.2A
• Output:
12V DC, 3.5A
External Battery Pack
• 12V LiMnO
2
disposable battery pack
• Capacity
when new, 10 hours of continuous operation
External Link
• UART port
• IrDA port