INSTRUCTIONS FOR USE
Z8A9F9 | IFU323EN00
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
60/60
Inconvenient
Cause
Solution
Note
The applanation is
not correctly
visualized during
the
measurement.
The tonometer is not
placed correctly in
respect of the
examined eye (F900).
Correctly place the
tonometer in
respect of the
examined eye.
Read the
indications given in
the instructions for
use.
The applanation is
not correctly
visualized during
the
measurement.
Missing fluorescein
instillation in the
examined eye.
Apply sodium
fluorescein as
indicated in the
instructions for use.
Always use
products compliant
with the current
legislation.
The applanation is
not correctly
visualized during
the
measurement.
The arm is bent.
Contact the
Technical
Assistance Center.
The device might
have been hit.