![CSO A900 Скачать руководство пользователя страница 30](http://html.mh-extra.com/html/cso/a900/a900_instructions-for-use-manual_2692899030.webp)
INSTRUCTIONS FOR USE
Z8A9F9 | IFU323EN00
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
28/60
Fig 22 - Positioning to align the measuring
prism with the microscope left shooting
channel (operator's left eye)
Fig 23 - Positioning to align the measuring
prism with the microscope shooting channel
(operator's right eye)
4.4
HOW TO PREPARE THE SLIT LAMP (Z800)
The microscope has to be adjusted so that the measuring prism
semicircles can be observed and clearly focused during the
measurement.
1
Rotate the arm with measuring prism support on the
observation axis of the microscope.
2
Make sure that the oculars of the slit lamp are adjusted to
compensate the operator's potential ametropia.
3
Adjust the device light intensity.
4
Inserting the blue filter in the slit lamp.
5
Open the slit diaphragm completely. The angle between the
light source and the microscope should be of 60° circa to
obtain a clear image without reflection.
4.5
HOW TO PREPARE THE SLIT LAMP (A900)
The microscope has to be adjusted so that the measuring prism
semicircles can be observed and clearly focused during the
measurement.