It is forbidden to use damaged measuring prisms. The contact
surface of the measuring prism must be inspected for contaminants
or damage (scratches, cracks and sharp edges) prior to every use.
Perform the check with a slit lamp microscope at 10/16
magnification.
Do not use the measuring prism beyond the date indicated indicated
on the production batch written on the package or after two years
from the date of first use.
During the measuring prism positioning use disposable gloves.