Z8A9F9 | IFU323EN00
INSTRUCTIONS FOR USE
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
29/60
1
Rotate the connection support to place the measuring prism
support on the observation axis of the microscope.
2
Make sure that the oculars of the slit lamp are adjusted to
compensate the operator's potential ametropia.
3
Adjust the device light intensity.
4
Inserting the blue filter in the slit lamp.
5
Open the slit diaphragm completely. The angle between the
light source and the microscope should be of 60° circa to
obtain a clear image without reflection.
4.6
HOW TO PREPARE THE SLIT LAMP (F900)
The microscope has to be adjusted so that the measuring prism
semicircles can be observed and clearly focused during the
measurement.
1
Make sure that the oculars of the slit lamp are adjusted to
compensate the operator's potential ametropia.
2
Adjust the device light intensity.
3
Inserting the blue filter in the slit lamp.
4
Open the slit diaphragm completely. The angle between the
light source and the microscope should be of 60° circa to
obtain a clear image without reflection.
4.7
HOW TO PLACE THE MEASURING PRISM
A wrong disinfection can cause cross-contamination between
patients and operator and damages to the measuring prism.
Detergent or disinfectant products residuals can cause eye
irritations.