INSTRUCTIONS FOR USE
Z8A9F9 | IFU323EN00
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
58/60
5.6
TROUBLESHOOTING
Inconvenient
Cause
Solution
Note
The device is not
stable during the
measurement or
during its
positioning.
The blocking screws
of the support are
loosened.
Fasten the screw of
the device support
with a suitable tool.
The device might
have been hit. If
necessary, contact
the Technical
Service for the
device calibration.
Wrong vertical
alignment of the
device in respect
to the
microscope.
The fastening screws
of the tonometer are
loosened (Z800).
Align the device
with the
microscope. Fasten
the device blocking
screws with a
suitable tool.
The device might
have been hit. If
necessary, contact
the Technical
Service for the
device calibration.
Wrong horizontal
alignment of the
device in respect
to the
microscope.
The fastening screws
of the tonometer are
loosened (Z800).
Align the device
with the
microscope. Fasten
the device blocking
screws with a
suitable tool.
The device might
have been hit. If
necessary, contact
the Technical
Service for the
device calibration.
Wrong horizontal
and vertical
alignment of the
device in respect
to the
microscope.
The arm is bent.
Contact the
Technical
Assistance Center.