5.3
MEASURING PRISM CLEANING
While cleaning the measuring prism use disposable gloves.
1
Carefully remove the measuring prism from the tonometer
support.
2
Wash it with cold running water for 30-60 seconds. If
necessary, use a neutral pH detergent (without irritating
agents) and a disposable cloth, soft which does not release lint.
Always dab in the same direction.
3
Rinse the measuring prism. Make sure that there is no
presence of detergent or contaminants. It there is still some
dirt repeat point 2.
Fig 34 - Measuring prism removal
Fig 35 - Measuring prism washing
4
Dry the measuring prism with a disposable cloth, soft which
does not release lint.
5
Check the absence of dirt on the measuring prism integrity.
6
Place the measuring prism in a clean and dry container.