HV Breakout Module – Mounting and installation
Installation Manual
22
Version 01.20 // 2021
4.4 Installing HV Breakout Modules
4.4.1 Before installation
WARNING!
A HV Breakout Module is applied in high-voltage applications.
Improper use can be life-threatening due to high voltage.
"
Make sure that this work is only carried out by qualified and trained
personnel.
"
Observe safety instructions.
WARNING!
The measurement module has to be connected to ground in order to
ensure user safety.
If the ground connection is not established, there is danger to life due to
high voltage.
"
Connect the measurement module to ground (PE).
"
Make sure that this work is only carried out by qualified and trained
personnel.
"
Observe safety instructions.
→ See chapter 4.4.2.4 "Ground connection".
WARNING!
The behavior of the CAN bus can be influenced by connecting a CAN bus
measurement module to an existing CAN bus system.
Improper use of a CAN bus system may cause life-threatening situations
and material damage.
"
Always connect CAN bus measurement modules to a separate CAN bus
system (measurement bus).
"
Make sure that this work is only carried out by qualified and trained
personnel.
NOTE!
i
The isolation barrier can be damaged due to aging, overvoltage, high
temperature and mechanical wear.
"
If there is any indication that the insulation barrier may be defective,
perform an insulation test immediately
and contact CSM. Do not put
the device in operation or continue to use it under any circumstances.