HV Breakout Module – How to use HV Breakout Modules
Installation Manual
30
Version 01.20 // 2021
5.2.3
Device configuration dialog
Fig. 5-3:
AutoConfig - XCP-Gateway
window, module connected via ECAT
"
Double-click on the device entry using the left mouse button.
⇨
The
Device configuration dialog
is displayed.
Fig. 5-4: Channel configuration dialog, module connected via ECAT
The number of available measurement channels is specified in the
Channels
selection menu
(left). The default settings are
3: U/I/P
(HV BM 1.x) and
9: U/I/P
(HV BM 3.1). The following
channels are active, depending on the settings:
f
3 or 9 channels:
U, I, P voltage, current, instantaneous power
f
2 or 6 channels:
U, I
voltage, current
f
1 or 3 channels:
U
voltage
f
1 or 3 channels:
I
current
This cannot be done directly on the CAN side. Proceed as follows in order to measure just
current on the CAN side as well:
"
Enable the option
Per channel configuration
in the
CAN
section (see below).
"
Then enter the CAN ID "0" or "0x0000" for the U and P channels in the
CAN ID
field of
the
Channel Configuration Dialog
.
The selection menu
Rate
(right) is used to set the measurement data rate which is valid for
all measurement channels.
NOTE!
i
The online-calculation of the measurement values for the instantaneous
power is always carried out at the highest rate of 1,000 kHz (HV BM 1.x) or
500 kHz (HV BM 3.1). The software filter is only applied after the instanta-
neous power has been calculated.
In the lower section of the dialog, the
shunt type that is installed
in the HV Breakout Module
is displayed.