5.2.1 Preparing configuration
"
Before starting an online configuration, make sure that
f
Measurement module and XCP-Gateway (or CAN interface) are properly connected
f
XCP-Gateway (or CAN interface) is properly connected to the data acquisition PC
f
CSMconfig is installed on the PC
f
the latest firmware version has been installed on the measurement module and the
XCP-Gateway
"
Start CSMconfig.
"
Select the required XCP-Gateway (or CAN interface) in the
Interface
dialog (
Options
|
Interface
or
Ctrl
+
I
) if it is not yet displayed in the status bar of CSMconfig.
5.2.2 Auto-Configuration
Auto-Configuration
provides the means to detect the measurement modules connected to
the bus and to read out configurations stored in a device. However, an automatic channel
configuration in its actual sense (e.g. setting the measurement range) is not performed. A
new configuration document is automatically created when executing the command.
Fig. 5-2:
Auto-Configuration
option
"
Select
Auto-Configuration (File | Auto-Configuration
or
Alt + A).
⇨
The bus is checked for available measurement modules.
⇨
The
AutoConfig - Tree View
window opens.
NOTE!
i
If
Auto-Configuration
fails to find any modules:
f
Make sure that the correct interface has been selected.
f
Check the connections and the power supply.
f
Is the module’s status LED lit?
f
Are the network settings correct?
f
Does the firewall block the Ethernet access from CSMconfig?
f
Is the CSMconfig installation up to date? Make sure that always the
latest version is installed.
"
If CSMconfig still doesn’t work, please contact our support.