IFU-000100 [Indice C] [SYMS I] [P.
14
/
18
]
Replacing the elastomer mat
Open the protective flap using the
rapid nut, turning it anti-clockwise.
Loosen the knurled nut and
remove the cooling jig.
Remove the ‘O’ ring weld.
Remove the adhesive PTFE
band covering the cooling jig.
Change the elastomer mat
supplied in the composition by
inserting it in the grove of the
cooling jig.
Cover the cooling jig and the
elastomer mat with the “G22”
adhesive PTFE band of size 45
mm x 30 mm supplied in the
composition (apply the adhesive
correctly).
Reinstall the ‘O’ ring weld.
Reinstall the cooling jig, then
check manually that the cooling
jig comes into contact with the
welding band, otherwise, check
the positioning of the end-stop
and adjust the cooling jig with
the help of the knurled nut.
Close the protective flap.