2.2.2 Installation safety rules
Installation
- To be safe, this equipment must be connected to an installation that is in conformity
(in France NFC 15-100), equipped with a ground connection, suitably protected
against ground faults.
- Operation on alternating current. The voltage and frequency must comply with the
indications given on the nameplate.
- In the event of damage to the mains cable, turn of the welder power supply
immediately. "Mortal danger!" Have the cable replaced by a specialist or contact
the after sales service at Cryo Bio System.
Warning
- If the equipment is used in a way that is not specified in the manual, this may have
an effect on the protection offered by the equipment.
- All modifications and repairs of an electrical nature must only be carried out by
qualified personnel, operating under the authority of the after-sales service
technician representing Cryo Bio System.
Descriptions
The SYMS I system consists of:
the SYMS I sealer,
a main power cable,
a set of spare parts
New: a control pedal is available as an option, under the reference 017171.
Energizing is carried out using the switch marked “U0” on the back panel and is
displayed by a green indicator light “D1” on the front panel.
The control pushbuttons or pedal will start the welding cycle.
Notes:
welding time (in seconds), factory setting: 2 s
press time (in seconds), factory setting: 4 s
The protection trap provides total safety against burns when the sealer is operating.
Similarly, the opening of the protection trap will cause the cycle to stop and block
the controls. However, the equipment will remain energized (indicated by the green
power indicator light).
A fuse protection covers any fault current that might be generated.