This document may not be reproduced, copied, transmitted or given to a party other than the end-user without the explicit
written authorization of Cryo Bio System.
IFU-000100 [Indice C] [SYMS I] [P.
11
/
18
]
4
ASSEMBLY AND ADAPTATION
Recommendation
: On exit from the factory, the SYMS I sealer is configured for use
with 0.3 and 0.5 ml CBS
TM
High Security straws and High Security Vitrification straws.
Sealing power
Cleaning
The welding power is pre-set in the factory
and is accessible on the potentiometer on
the rear face.
This setting is specific to the machine.
It is strongly advised not to modify this
setting.
However, and only on the advice of a
maintenance technician, the user may be
called upon to modify this setting if the
sealing is judged inadequate.
The removable parts are:
the cooling jig,
the welding electrode assembly.
They may be removed without difficulty and
decontaminated, provided that there is:
- no immersion.
Decontamination can be carried out with:
- Sanytex diluted to 5% or another equivalent
product,
- a phase of rinsing with distilled water,
- drying with alcohol.
Open the protective flap using rapid nut,
turning it anti-clockwise.
Loosen the knurled nut and remove the
cooling jig.
Remove the welding electrode assembly
- Remove the stop and straw guide by
pulling them out of the electrode assembly.