IFU-000100 [Indice C] [SYMS I] [P.
13
/
18
]
Changing the welding tape band
Changing the cooling jig band
Remove the stop and straw guide by pulling
them out of the electrode assembly.
Remove the adhesive PTFE band covering
the sealing tape.
Cover the sealing tape with the PTFE
adhesive “G9” supplied in the composition.
Make sure that the adhesive is correctly
applied.
Reinstall the end-stop and the straw guide.
Remove the ‘O’ ring weld.
Remove the adhesive PTFE
band covering the cooling jig.
Cover the cooling jig with the
“G22” adhesive PTFE band size
45mm x 30mm supplied in the
composition.
Make sure that the adhesive is
correctly applied.
Reinstall the ‘O’ ring weld.
Reassemble the equipment
and check manually that the
cooling jig comes into contact
to the welding band, otherwise
check the positioning of the
end stop and adjust the cooling
jig with the help of the knurled
nut.