18
]
5
PROCEDURE
5.1 Operation of the individual straw sealer
The sealer does not require any heating up time. It can be used without any problem
in a laminar flow hood or a controlled temperature environment, because the
thermal pulse welding principle limits the release of calories.
However, after a series of 10 seals, it is recommended to comply with a rest time
of five minutes in order to obtain optimum operation.
Connection
Sélection chauffe
Connect the mains power cable “Cb1” to
the socket in the input filter “U0” identified by:
POWER SUPPLY
For use with the control pedal (available as
an option), connect the cable of the control
pedal to the socket “J0” identified by:
FOOT SWITCH
– Set the “U0” main switch to the ‘ON’ position.
– Check that the green “D1” Power On indicator is lit.