The following screen appears after selecting “SUPPLY VOLTAGE TEST”:
If the screen doesn’t appear, one of the internal supply voltages are off; switch the device off and
on again and try again.
The following screen appears after selecting “SUPPLY VOLTAGE TEST”:
The following screen appears after a waiting period of 10 s.
Note: If the current source test is without error and the device nevertheless works ok, please
change the current source fuse on the back panel.
Please read chaper “safety instructions”
Fuse: Super quick acting 10A fuse 6,3*32mm, 600VAC, 50000A breaking capacity (or greater)
RS components #209-9383 (Germany)
Use only this fuse.