DO6 Handbook
Jan 2006
93 of 107 pages
8.14.15 SENSe:AVERage:COUNt
DESCRIPTION:
Sets the number of measurements to be used for calculating the mean
resistance.
SYNTAX:
SENSe:AVERage:COUNt P1
Meaning of parameter Pn
QUERY FORM:
SENSe:AVERage:COUNt?
RESPONSE:
A1
Meaning of parameter An
NOTE:
Command not allowed in calibration mode.
Command not allowed when measurement is running.
8.14.16 SENSe:CORRection:OFFset
DESCRIPTION:
Start zero-offset measurement for automatic thermal-EMF
compensation disabled (“MAN ZERO”).
SYNTAX:
SENSe:CORRection:OFFset?
No parameter
QUERY FORM:
No query form
NOTE:
Command not allowed in calibration mode.
Command not allowed when measurement is running.