6.5.4.3 Alternating measurement mode
Alternating measurement mode means that the test current is switched on when the START
button is pressed and not switched off until the STOP button is pressed. The current source is
switched on and off continuously during the measurement to suppress any thermal EMF’s, so
that the instrument remains permanently correctly “zeroed”. Select this measurement mode
for ultra precise measurements that are not time critical.
Mean values of n values are displayed. One digitization takes about 2 s (Z1, N=1). While the
measurement is in progress, the animation (-) indicator displayed on the lower left flashes at
second intervals to show that the measurement is running. This setting cannot be used in
conjunction with time constants Z2, Z3 or with an inductive load.
The setting MAN ZERO / AUTOZERO is ignored.
Pressing CHANGE displays the following screen:
The cursor sits over the first zero. Pressing
↑↓
increases or decreases the numerical value,
while
→
moves the cursor one position to the right within the input field. After selecting the
value, pressing ENTER saves the value and closes the menu.