6.5.4.4 Cooling curve
The cooling curve measurement mode is allowed in conjunction with all times constants, and
manual and automatic zero offset.
It is not allowed, however, in conjunction with comparator, automatic measuring range and
automatic temperature compensation. The settling OHM/m is also ignored. Nor in this case
is it possible to change the measuring range during the measurement time constant Z1.
Pressing CHANGE displays the following screen:
The cursor sits over the first zero. Pressing
↑↓
increases or decreases the numerical value,
while
→
moves the cursor one position to the right within the input field. After selecting the
value, pressing ENTER saves the value and closes the menu.
The INTERVAL TIME is the time between two measurements; it must always be shorter than
the END TIME.
The END TIME is the time at which the measurement is terminated. Shown later as MAX in
the display. It must always be greater than the INTERVAL TIME, which is the time between
two measurements.