NOTE:
Command not allowed in calibration mode.
Command not allowed when measurement is running.
8.5.4 CALCulate:LIMit:RESet
DESCRIPTION:
Behaviour of comparator function. The comparator is reset with Start
measurement (static behaviour) or not reset (dynamic behaviour).
SYNTAX:
CALCulate:LIMit:RESet P1
Meaning of parameter Pn
QUERY FORM:
CALCulate:LIMit:RESet ?
RESPONSE:
A1
Meaning of response An
NOTE:
Command not allowed in calibration mode.
Command not allowed when measurement is running.
8.5.5 CALCulate:LIMit:LOWer
DESCRIPTION:
Sets the lower comparator limit. This value is not adopted, however,
until the CALCulate:LIMit:ACKNowledge? command is received, once the upper
comparator limit has also been transferred using the CALCulate:LIMit:UPPer command.
SYNTAX:
CALCulate:LIMit:LOWer P1
Meaning of parameter Pn