5. Operating instructions in brief
After switching on the instrument, the operating language can be selected in the identification
menu.
Pressing both arrow buttons simultaneously opens the configuration program. ENTER confirms
the selected menu option. ESC can be used to return from any option in the configuration menu
back to the next menu option down. If a value needs to be changed e.g. limit, arrows appear
above the START/STOP buttons to move the cursor to the left/right. The numerical value is
changed using the up/down arrow buttons (on the right hand side) on the front panel.
Meas. range
Automatic range 1…4
20.000 m
Ω
200.00
m
Ω
2.0000
Ω
20.000
Ω
200.00
Ω
2.0000 k
Ω
20.000 k
Ω
200.00 k
Ω
The limitation of the autorange mode is
only possible with the PC software. With
this limitation you get in the autorange
mode a higher measurement speed. (e.g.
range 1…4)
Limit value
Upper limit
Lower limit
xxx.xxxx y
Ω
xxx.xxxx y
Ω
Load
Resistive load: Z1 (little time constant)
Inductive load: Z2 (middle time constant)
Inductive load: Z3 (high time constant)
Meas. mode
Continuous
Single shot
Alternating
Cooling curve
Average of xx values
xxx value after start
Average of xx values
Interval time xxxx s
End time xxxx s
Temp. comp.
Off
Copper
930 ppm
Aluminium
4030 ppm
Brass 63
1500 ppm
Brass 80
1600 ppm
Tungsten
4400 ppm
Nickel
6180 ppm
Platinum
3900 ppm
8 further materials with the
belonging TCs you can add
only with the PC software
Zero mode
Auto-Zero
Man-Zero
Содержание DO6
Страница 7: ...DO6 Handbook Jan 2006 7 of 107 pages 2 5 Block diagram...
Страница 56: ...DO6 Handbook Jan 2006 56 of 107 pages 8 2 SPCI registers...
Страница 100: ...DO6 Handbook Jan 2006 100 of 107 pages...
Страница 101: ...DO6 Handbook Jan 2006 101 of 107 pages 8 16 2 Communication using Fast selection...
Страница 102: ...DO6 Handbook Jan 2006 102 of 107 pages...