DEPRESSURIZe ?
st : yes Can : no
2
Some situations may cause interruption and automatic cancellation of the cycle when:
the ideal pressure/temperature is not reached due to steam/pressure leak, lack of water or overloaded
chamber. In this case, the cancellation will occur within 35 minutes at the most;
there’s a p
ower outage or voltage fluctuations.
Note:
The
depressurization will occur when power returns;
t
urning the autoclave on, there’s already some pressure in the chamber. When that happens, the autoclave
will beep continuously, the display will show
CYCLE cancelled
and the autoclave will depressurize;
Confirm the cancellation of the cycle manually by pressing the
CANCEL
key;
Check the possible causes, take the necessary measures to correct the problem and perform a new
cycle to reprocess the articles according to the instructions on “
How to Use the Autoclave Vitale Plus
”
(Page 13). Before starting a new cycle, t
he operator must check if there is water left in the chamber,
which must be removed manually through the door with a clean and dry cloth that does not shed.
Attention!
For your safety, remember to use PPE (Personal Protective Equipment, like proper latex gloves);
Note:
In the situations
b
and
c
above, if there’s pressure inside the chamber, the display will show the question:
By pressing the
START
key, the valve opens for depressurization; by pressing the
CANCEL
key,
the depressurization will happen with the valve closed, this will prevent the liquid from boiling
over inside the containers when the liquids sterilization cycle was selected;
If the operator needs to interrupt/cancel the heating, sterilization or drying cycle, just press the
CANCEL
key. In
this case, after the beep and the indication
cycle cancelled
on the display, the autoclave
will beep
2
continuously until the pressure falls to 0,0 kgf/cm and the
CANCEL
key is pressed again to confirm the
cancellation of the cycle (there will be no effect in case the
CANCEL
key is pressed while there’s still pressure in
the chamber);
Cancellation of the Cycle -
If the cycle is cancelled, for any reason, the display will alternate between two
messages, the message indicating the cycle cancellation (
CYCLE CANCELLED P:0.00 kgf/cm2
) and the message
indicating the possible cause for that, which can be:
heating time limit
- the heating time limit was reached and there was no sterilization;
incompatible Time x PRESSURE
- The relation between the temperature and pressure of the chamber is incoherent;
subPRESSure DETECTED
- During the sterilization, the pressure decreased below the acceptable level;
OverPRESSure DETECTED
- During the sterilization, the pressure increased above the acceptable level;
CANCELLED by user
- The operator pressed the
CANCEL
key;
no power
- There was a power outage at the workplace (during the sterilization stage only).
In case of activation of one of the safety devices (sudden steam escape), generally caused by obstruction of
the internal steam exit or by an obstruction of the solenoid valve, wait for complete depressurization before
opening the door. The display will show
cycle cancelled
.
ADVERSE SITUATIONS
a
b
1
c
3
17
4