I N S T R U C T I O N    M A N U A L

DEPRESSURIZe ?

st : yes   Can : no

2

 

Some situations may cause interruption and automatic cancellation of the cycle when:

 

 

the  ideal  pressure/temperature  is  not  reached  due  to  steam/pressure  leak,  lack  of  water  or  overloaded 
chamber. In this case, the cancellation will occur within 35 minutes at the most;

 

there’s a p

ower outage or voltage fluctuations. 

Note:

 The 

depressurization will occur when power returns;

 

 

 

t

urning the autoclave on, there’s already some pressure in the chamber. When that happens, the autoclave 

will beep continuously, the display will show 

CYCLE cancelled

 and the autoclave will depressurize;

 

 

 

  Confirm the cancellation of the cycle manually by pressing the 

CANCEL

 key;

 

 

 

 

Check  the  possible  causes,  take  the  necessary  measures  to  correct  the  problem  and  perform  a  new 

cycle  to  reprocess  the  articles  according  to  the  instructions  on  “

How  to  Use  the  Autoclave  Vitale  Plus

” 

(Page  13).  Before  starting  a  new  cycle,  t

he  operator  must  check  if  there  is  water  left  in  the  chamber, 

which  must  be  removed  manually  through  the  door  with  a  clean  and  dry  cloth  that  does  not  shed. 

Attention!

 For your safety, remember to use PPE (Personal Protective Equipment, like proper latex gloves);

 

Note:

 In the situations 

b

 and 

c

 above, if there’s pressure inside the chamber, the display will show the question:

 

 

   

 

By pressing the 

START

 key, the valve opens for depressurization; by pressing the 

CANCEL

 key, 

the depressurization will happen with the valve closed, this will prevent the liquid from boiling 
over inside the containers when the liquids sterilization cycle was selected;

 

 

If the operator needs to interrupt/cancel the heating, sterilization or drying cycle, just press the 

CANCEL

 key. In 

this  case,  after  the  beep  and  the  indication 

cycle  cancelled

  on  the  display,  the  autoclave 

will  beep 

2

continuously  until  the  pressure  falls  to  0,0  kgf/cm   and  the 

CANCEL

  key  is  pressed  again  to  confirm  the 

cancellation of the cycle (there will be no effect in case the 

CANCEL

 key is pressed while there’s still pressure in 

the chamber);

 

 

Cancellation  of  the  Cycle  -

  If  the  cycle  is  cancelled,  for  any  reason,  the  display  will  alternate  between  two 

messages,  the  message  indicating  the  cycle  cancellation  (

CYCLE  CANCELLED  P:0.00  kgf/cm2

)  and  the  message 

indicating the possible cause for that, which can be:

 

 

heating time limit

 - the heating time limit was reached and there was no sterilization;

 

 

incompatible Time x PRESSURE

 - The relation between the temperature and pressure of the chamber is incoherent;

 

 

subPRESSure DETECTED

 - During the sterilization, the pressure decreased below the acceptable level;

 

 

OverPRESSure DETECTED

 - During the sterilization, the pressure increased above the acceptable level;

 

 

CANCELLED by user

 - The operator pressed the 

CANCEL

 key;

 

 

no power

 - There was a power outage at the workplace (during the sterilization stage only).

 

 

In case of activation of one of the safety devices (sudden steam escape), generally caused by obstruction of 
the  internal  steam  exit  or  by  an  obstruction  of  the  solenoid  valve,  wait  for  complete  depressurization  before 
opening the door. The display will show 

cycle cancelled

.

ADVERSE SITUATIONS

a

b

1

c

3

17

4

Содержание Vitale Plus 12

Страница 1: ...uce its environmental impacts and continuous training of its employees achieving this way a sustainable profitability and the maximization of the company s value Rev 2 Develop innovative solutions to...

Страница 2: ......

Страница 3: ...rist foli s Biosafety Consultant who provides a valuable assistance in the coordination of Crist foli s Biosafety Project complementary products research training of our employees representatives and...

Страница 4: ...ymbols Introduction Important Safety Information Installation Instructions Autoclave Components Identification Safety Devices Cycles and Messages Safety Notes How to Use the Autoclave Vitale Plus Alti...

Страница 5: ...nding Protection Terminal Fusible Autoclavable Date of manufacture Boas Pr ticas de Fabrica o 4 P B F LOT SN Fragile Handle with care Dangerous Electrical Tension Serial Number Batch code Keep away fr...

Страница 6: ...ation we recommend extra attention to this manual Autoclave sterilization although quicker and more efficient requires totally different procedures and handling while still keeping a simple operation...

Страница 7: ...y kind of use of this equipment other than the ones described in this manual SPECIAL MEASURES AND OBSERVATIONS DURING THE USE OF THE AUTOCLAVE Always make sure the door handle is properly locked befor...

Страница 8: ...he use of an exclusive circuit breaker for the outlet where you intend to install your autoclave If after following all the specifications the electrical current is still oscillating contact your loca...

Страница 9: ...of the external steam exit 8 Fig 1A Fig 3 WATER V CUT DO NOT USE THE LID KEEP THE HOSE ABOVE THE WATER LEVEL 5 liters 1 3 gallons Table 2 X NEVER CUT THE CENTRAL PIN OFF Fig 1 EXTERNAL STEAM EXIT VSPF...

Страница 10: ...use interference in the cycle EXTERNAL STEAM EXIT Located in the back part of the autoclave Fig 1 page 8 it has a diameter of 5 16 for the connection of the discharge hose which is then connected to t...

Страница 11: ...7 Door Gasket Fig 5 Display Start Key Mode Key Cancel Key Lid Panel Keyboard Handle Fig 4 CRIST FOLI A U T O C L A V E S I N S T R U C T I O N M A N U A L Fig 8 Fig 10 Fig 9 250 ml 50 ml 200 ml 150 m...

Страница 12: ...onents Identification Figs 6 and 7 page 10 VSPF VALVE Valve that works as an anti vacuum system patent pending FUSE Safety device which purpose is to protect the building electrical wiring against pea...

Страница 13: ...an no perform cleaning see manual perform biological test cooling 00 00 T 000 C p 0 00 cooling 00 00 cooling T 000 C power outage subpressure detected cancelled by user overpressure detected temp x pr...

Страница 14: ...ymbol 14 appears in some places on the autoclave This means those items require special attention and that the user must observe their references in the Instruction Manual provided with the equipment...

Страница 15: ...re for the new cycle is reached the autoclave will resume operation showing heating up When the solenoid valve shuts down the display will show the temperature and pressure rising gradually When the a...

Страница 16: ...the CANCEL key to return to the program selection mode ATTENTION Even after the beeps that indicate the conclusion of the cycle the contents of the chamber will still be hot Never touch the internal...

Страница 17: ...me ALTITUDE CALIBRATION In order for the autoclave to work properly it is necessary to perform the altitude calibration procedure Calibrate the altitude of the autoclave as shown below for this proced...

Страница 18: ...valve closed this will prevent the liquid from boiling over inside the containers when the liquids sterilization cycle was selected If the operator needs to interrupt cancel the heating sterilization...

Страница 19: ...ature of drained water Total heat in joules transmitted in one hour Proper working temperature range Proper working altitude 12 liters Aluminum 19 4 kg including components Stainless steel 21 2 kg inc...

Страница 20: ...eplace the air filter of the VSPF valve monthly Replace the thermal paste of the heating element annually The component lid Item 2 page 9 must be replaced each 10 years Replace the internal hoses annu...

Страница 21: ...hamber Preventive cleaning stainless steel chamber Door gasket cleaning Door gasket replacement Preventive Maintenance of the VSPF valve and replacement of the internal safety valve Air filter replace...

Страница 22: ...pose of the label located in the back of the equipment is to identify the autoclave technical data ATTENTION The removal of the identification label and or any other labels stickers affixed to the pro...

Страница 23: ...on See How to Proceed when Service is Needed Page 24 THE AUTOCLAVE DOES NOT SWITCH ON POSSIBLE CAUSES SOLUTION The operator connected the autoclave to the power source but didn t press the Start key N...

Страница 24: ...g the sterilization of 26 envelopes at a time 10 x 23 cm containing 6 instruments each Remember to leave some space between the envelopes to allow good steam circulation and optimize drying THE AUTOCL...

Страница 25: ...ive Maintenance Item 7 Page 19 Crist foli Equipamentos de Biosseguran a Ltda makes clear its concern already demonstrated by the implementation of the Environmental Management System according to stan...

Страница 26: ...rcuits fire etc Damage caused by humidity excessive exposure to sun light and salinity Sinister theft or robbery Improper use installation of the equipment and or plugging it into an incorrect voltage...

Отзывы: