ATTENTION!

 For any replacement of parts, contact your local dealer or the authorized technical assistance office. It’s 

strongly  not  recommended

  the  replacement  of 

any 

parts  by  non-qualified  people.  We  have  listed  below  the  most 

frequent problems and possible solutions the operator can try in his/her own office:

TROUBLESHOOTING

POSSIBLE CAUSES

SOLUTION

• No power ---------------------------------------------------------------------------
•The autoclave is not connected to the outlet ------------------------------
• The 

Start

 key was not pressed to turn the equipment on ---------------

• The fuse has burned out ---------------------------------------------------------

• Defective electronic circuit ----------------------------------------------------

•·Check if there’s a power outage in your area/building;
•·Connect the power cable to the electrical outlet;
•·Press  the 

Start

  key,  the  initial  information  (brand,  model  and 

software version) is shown on the display ;

•·Replace the fuse next to the power connector. In order to replace 
the  fuse,  unplug  the  equipment,  remove  the  fuse  holder  cap, 
replace  the  defective  fuse,  screw  the  fuse  holder  cap  back  into 
place and turn the equipment on;
•·See “

How to Proceed when Service is Needed

” (Page 24);

THE AUTOCLAVE DOES NOT SWITCH ON

POSSIBLE CAUSES

SOLUTION

• The operator connected the autoclave to the power source but 

didn’t press the 

Start

 key -------------------------------------------------------

• Nothing happens when the 

Start 

key is pressed --------------------------

• The heating element has burned out ----------------------------------------
• Defective thermostat ------------------------------------------------------------
• Defective electronic circuit ----------------------------------------------------

•·Press the 

Start

 key after selecting the cycle on the 

Mode

 key;

•·See “

How to Use the Autoclave Vitale Plus

” (Page 13).

•·See “

How to Proceed when Service is Needed

” (Page 24);

THE AUTOCLAVE SWITCHES ON BUT DOES NOT HEAT UP

If the problem persists after the verification of all the items listed, contact your local dealer.

POSSIBLE CAUSES

SOLUTION

•  Equipment  installed  where  the  altitude  is  above  the 

recommended ------------------------------------------------------------------

•  In  order  for  the  autoclave  to  work  properly,  it  is  necessary  to 
perform the altitude calibration procedure. Calibrate the altitude of 
the autoclave as shown on topic “

Altitude Calibration

” (Page 16).

THE AUTOCLAVE SWITCHES ON, HEATS UP BUT THERE’S NO PRESSURIZATION

• See “

How to Proceed when Service is Needed

” (Page 24);

• Defective temperature sensor ------------------------------------------------

22

CRISTÓFOLI

A U T O C L

A V E S

I N S T R U C T I O N    M A N U A L

Содержание Vitale Plus 12

Страница 1: ...uce its environmental impacts and continuous training of its employees achieving this way a sustainable profitability and the maximization of the company s value Rev 2 Develop innovative solutions to...

Страница 2: ......

Страница 3: ...rist foli s Biosafety Consultant who provides a valuable assistance in the coordination of Crist foli s Biosafety Project complementary products research training of our employees representatives and...

Страница 4: ...ymbols Introduction Important Safety Information Installation Instructions Autoclave Components Identification Safety Devices Cycles and Messages Safety Notes How to Use the Autoclave Vitale Plus Alti...

Страница 5: ...nding Protection Terminal Fusible Autoclavable Date of manufacture Boas Pr ticas de Fabrica o 4 P B F LOT SN Fragile Handle with care Dangerous Electrical Tension Serial Number Batch code Keep away fr...

Страница 6: ...ation we recommend extra attention to this manual Autoclave sterilization although quicker and more efficient requires totally different procedures and handling while still keeping a simple operation...

Страница 7: ...y kind of use of this equipment other than the ones described in this manual SPECIAL MEASURES AND OBSERVATIONS DURING THE USE OF THE AUTOCLAVE Always make sure the door handle is properly locked befor...

Страница 8: ...he use of an exclusive circuit breaker for the outlet where you intend to install your autoclave If after following all the specifications the electrical current is still oscillating contact your loca...

Страница 9: ...of the external steam exit 8 Fig 1A Fig 3 WATER V CUT DO NOT USE THE LID KEEP THE HOSE ABOVE THE WATER LEVEL 5 liters 1 3 gallons Table 2 X NEVER CUT THE CENTRAL PIN OFF Fig 1 EXTERNAL STEAM EXIT VSPF...

Страница 10: ...use interference in the cycle EXTERNAL STEAM EXIT Located in the back part of the autoclave Fig 1 page 8 it has a diameter of 5 16 for the connection of the discharge hose which is then connected to t...

Страница 11: ...7 Door Gasket Fig 5 Display Start Key Mode Key Cancel Key Lid Panel Keyboard Handle Fig 4 CRIST FOLI A U T O C L A V E S I N S T R U C T I O N M A N U A L Fig 8 Fig 10 Fig 9 250 ml 50 ml 200 ml 150 m...

Страница 12: ...onents Identification Figs 6 and 7 page 10 VSPF VALVE Valve that works as an anti vacuum system patent pending FUSE Safety device which purpose is to protect the building electrical wiring against pea...

Страница 13: ...an no perform cleaning see manual perform biological test cooling 00 00 T 000 C p 0 00 cooling 00 00 cooling T 000 C power outage subpressure detected cancelled by user overpressure detected temp x pr...

Страница 14: ...ymbol 14 appears in some places on the autoclave This means those items require special attention and that the user must observe their references in the Instruction Manual provided with the equipment...

Страница 15: ...re for the new cycle is reached the autoclave will resume operation showing heating up When the solenoid valve shuts down the display will show the temperature and pressure rising gradually When the a...

Страница 16: ...the CANCEL key to return to the program selection mode ATTENTION Even after the beeps that indicate the conclusion of the cycle the contents of the chamber will still be hot Never touch the internal...

Страница 17: ...me ALTITUDE CALIBRATION In order for the autoclave to work properly it is necessary to perform the altitude calibration procedure Calibrate the altitude of the autoclave as shown below for this proced...

Страница 18: ...valve closed this will prevent the liquid from boiling over inside the containers when the liquids sterilization cycle was selected If the operator needs to interrupt cancel the heating sterilization...

Страница 19: ...ature of drained water Total heat in joules transmitted in one hour Proper working temperature range Proper working altitude 12 liters Aluminum 19 4 kg including components Stainless steel 21 2 kg inc...

Страница 20: ...eplace the air filter of the VSPF valve monthly Replace the thermal paste of the heating element annually The component lid Item 2 page 9 must be replaced each 10 years Replace the internal hoses annu...

Страница 21: ...hamber Preventive cleaning stainless steel chamber Door gasket cleaning Door gasket replacement Preventive Maintenance of the VSPF valve and replacement of the internal safety valve Air filter replace...

Страница 22: ...pose of the label located in the back of the equipment is to identify the autoclave technical data ATTENTION The removal of the identification label and or any other labels stickers affixed to the pro...

Страница 23: ...on See How to Proceed when Service is Needed Page 24 THE AUTOCLAVE DOES NOT SWITCH ON POSSIBLE CAUSES SOLUTION The operator connected the autoclave to the power source but didn t press the Start key N...

Страница 24: ...g the sterilization of 26 envelopes at a time 10 x 23 cm containing 6 instruments each Remember to leave some space between the envelopes to allow good steam circulation and optimize drying THE AUTOCL...

Страница 25: ...ive Maintenance Item 7 Page 19 Crist foli Equipamentos de Biosseguran a Ltda makes clear its concern already demonstrated by the implementation of the Environmental Management System according to stan...

Страница 26: ...rcuits fire etc Damage caused by humidity excessive exposure to sun light and salinity Sinister theft or robbery Improper use installation of the equipment and or plugging it into an incorrect voltage...

Отзывы: