The autoclaves
Vitale Plus
have the following safety devices:
TEMPERATURE X PRESSURE DATA CROSSING ELECTRONIC SYSTEM -
Internal system of the equipment which will check
the cycle, in case any problem is detected while reading the pressure in the chamber or if the autoclave pressure
exceeds the safety limit, the cycle will be cancelled automatically.
2
DOOR GASKET -
In case the pressure goes beyond 3 kgf/cm or 294 kPa, the door gasket will detach from the edge
of the door making a loud noise. See “
Autoclave Components Identification”
(Figs. 6 and 7, page 10).
VSPF VALVE -
Valve that works as an anti-vacuum system (patent pending).
FUSE -
Safety device which purpose is to protect the building electrical wiring against peaks of energy. The fuse used
is the 20 Glass AGLF- Quick Action. In case the operator wishes to replace the fuse personally, the table below will
provide the necessary information;
SAFETY DEVICES
1
2
3
4
5
6
Table 3
THERMOSTAT
- Internal safety device of the equipment. Its function is to limit the excessive heating of the chamber
during the sterilization cycles or in case of a circuit board malfunction.
POWER CONTROL ELECTRONIC SYSTEM
- Internal system of the equipment which monitors the temperature and the
pressure of the autoclave throughout the operation.
VOLTAGEM
FUSE
(Vitale Plus 12)
127 V (114 V - 140 V)
127 V
10 A (250 V)
FUSE
(Vitale Plus 21)
12 A (250 V)
220 V
220 V (198 V - 253 V)
6 A (250 V)
8 A (250 V)
AC VOLTAGE LINE