12 13
Measuring Range
Accuracy
Memory
Power supply
Pressure: 30~280 mmHg
Pulse: 40~199 Beat/min
Static Pressure: ± 3mmHg Pulse: ± 5%
90 Memories
2x 1.5V Batteries (LR03 or AAA)
use alkaline battery, measure above 200 times
Operating condition:
+5
40
.
15%RH~93%RH
Atmospheric pressure: 70kPa~106kPa
℃ ~+ ℃
Storage condition:
-20 ~+55
. 0%RH~93%RH
Atmospheric pressure:50kPa~106kPa
℃
℃
Dimensions
Weight:
Classification
Wrist cuff
Approx.: 62(W)X78(H)X31(D)mm
Approx.: 130g, excluding batteries
Type BF
13,5 ~19,5 cm
Specifications may be changed without notice in the event of improvement
being made.
1.Type of protection against electric shock: INTERNALLY POWERED EQUIPMENT.
2.Degree or protection against electric shock: TYPE BF APPLIED PART.
3.Mode of operation: CONTINUOUS OPERATION.
4.Equipment not suitable for category AP&APG equipment use in presence.
STATEMENT
The system might not meet its performance specifications if stored or used
outside the temperature and humidity as mentioned below:
Operating conditions: +5
℃
~+40
℃
15%RH~93%RH 70kPa~106kPa
Storage conditions: -20
℃
~+55
℃
.
0%RH~93%RH
.
TROUBLESHOOTING
ERROR DISPLAY
POSSIBLE CAUSE
HOW TO CORRECT
Nothing is displayed
when you push the
POWER button or the
battery icon flashes
lf you have trouble in using the unit please check the following points first.
No battery installation
Battery worn out
The polarities of batteries
placed wrongly
Insert batteries
Replace new batteries
Insert battery in the
correct polarities
E2E4: have shaking
while measurement
Battery icon on
The systolic pressure
Value or diastolic
Pressure value too
high
The systolic pressure
Value or diastolic
Pressure value too low
Hand or body shaking
while measurement
Battery low power
1.The arm cuff was held
lower than your heart
2.The arm cuff was not
attached properly
3.You moved your body or
spoke during measurement
1.The arm cuff was held
higher than your heart
2.you moved your body or
Spoke during measurement
keeping static and correct
gesture to measure again
Replace battery and
measure again
keeping correct
position and gesture
to measure again
E1: can’t normally
Increase pressure
Check your cuff if any
air leakage
Replace the cuff with
new one
E3: inflate pressure too
high
Re-measure or send
back to the dealer for
re-calibrate pressure
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Group 1
Class B
N
.
A
.
N
.
A
.
Guidance and manufacturer’s declaration – electromagnetic emissions
Emissions
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Appendix 1 Guidance and Manufacturer Declaration Tables
The Model BPM220 Series Electronic Blood Pressure Monitor is intended for
use in the electromagnetic environment specified below. The customer
or the user of the Model BPM220 Series Electronic Blood Pressure Monitor
should assure that it is used in such an environment.
The
Series Electronic Blood
Pressure Monitor uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
Model BPM220
The
Series Electronic Blood
Pressure Monitor is used in home and it’s
powered by DC 3V
Model BPM220