background image

11

10

3.  Place a new storage bottle in the biosafety cabinet and unbag.
4.  Disconnect the filled waste bottle on the right side of the unit and replace 

with the new storage bottle for cell collection (harvest).

5.  Connect the power cord, turn the MicroDEN

 

unit “ON”, and select “Harvest” 

using the function button.

6.  Differentiation medium remaining in the left side inlet bottle will perfuse 

into the flasks.

7.  As medium is perfusing, carefully remove the flasks from the thermoform tray 

and rotate so that the outlet port is down and the inlet port is up (Figure 5). 

NOTE:

 It may be necessary to remove the tubing from the right side tubing 

holder.

8.  Allow perfusion to continue when the inlet bottle is empty of differentiation 

medium. Air will perfuse into the flasks and medium will drain into the 
harvest bottle.

9.  Remove the flasks from the thermoform tray and slide the flasks side-to-side 

as medium drains from the flasks into the harvest bottle (Figure 6). 

The sliding motion causes turbulence that allows the cultured cells to 
detach and suspend into solution.

Corner notch 
on flask in 
lower left 
position

3.  Place a new storage bottle in the biosafety cabinet and unbag.
4.  Open the bottle. 

To unscrew the cap, turn the bottle while keeping the cap and tubing set 
in place so the tubes don’t twist (Figure 1). Take care not to touch the dip 
tubes to any non-sterile surface.

5.  Add 80.0 mL differentiation medium to the new bottle.
6.  Insert tubing set and place the cap back onto the bottle. While pressing 

down on the tubing set, rotate the orange cap to tighten the assembly to the 
bottle (Figure 1). Place the bottle back into the thermoform tray.

7.  Disconnect the empty bottle and replace with the new storage bottle 

containing differentiation medium.

Disconnect the empty bottle MPCs and connect the new storage bottle 
MPCs to the inlet side tubing.

8.  Transfer the Corning® MicroDEN®

 

system from the biosafety cabinet into the 

incubator.

9.  Connect the power cord, turn the unit “ON”, and select “Culture” using the 

function button on the unit.

A timer will display elapsed time when culture begins; the pumps will then 
perfuse differentiation medium at 8.0 µL/min. into each flask.

Imaging 

The flasks can be imaged using most standard microscopes. The flasks do not 
need to be removed from the consumable assembly for imaging. If the entire 
consumable assembly does not fit in the microscope for imaging, the flasks can 
be removed from the thermoform tray. Take care not to accidentally disconnect 
the tubing outside a sterile environment. Alternatively, the flask can be dis-
connected from the tubing for imaging. To maintain sterility, sterile filters or 
Luer Lock caps should be placed on the inlet and outlet ports of the flasks. 
The tubing can also be capped using sterile filters or Luer Lock caps. Be sure to 
employ proper aseptic techniques.

9. Harvest (Day 6)

The Corning MicroDEN

 

unit has a “Harvest” function that perfuses cold buffer 

from the inlet bottle to remove cells from the flask. A new storage bottle can 
be used for both buffer (inlet) and cell collection (outlet) by disconnecting the 
MPCs and replacing with a new bottle.

1.  Turn the MicroDEN

 

unit “OFF”, disconnect the power cord, and remove from 

the incubator.

2.  Lightly spray the MicroDEN

 

system with 70% ethanol and place inside the 

biosafety cabinet.

Do not spray ethanol directly into the filters as wetting the filters will 
cause them to restrict gas exchange, and the bottles can pressurize as the 
buffer is perfused.

Outlet ports 
are down

Inlet ports 
are up

Figure 5. 

Set up for harvest (Day 6). 

Remove tubing 
from the tubing 
holder

Содержание MicroDEN

Страница 1: ...hnical information visit www corning com lifesciences or call 800 492 1110 Outside the United States call 1 978 442 2200 or contact your local Corning sales office Corning Incorporated Life Sciences 8...

Страница 2: ...autoclave this unit Do not perform the sterilization cycle of an incubator with the MicroDEN unit inside Do not use acetone to clean the unit Always use the provided external power supply to power the...

Страница 3: ...nts 1 Corning MicroDEN unit Cat No 431276 requires standard electrical outlet 2 Corning consumable set Cat No 431277 3 Corning storage bottle Cat No 431278 4 Enriched monocytes MOs 5 Differentiation m...

Страница 4: ...t a wall to prevent disturbing the cultured cells The cell culture flasks may be seeded with either enriched MOs or with peripheral blood mononuclear cells PBMCs where the MOs are enriched through pla...

Страница 5: ...ow through the unclamped tube Method 2 Pump Perfusion 1 Unclamp the pumps by pressing upwards at the dotted triangle on the pump head 2 Place the tubing onto the pump rollers The colored tubing marker...

Страница 6: ...emptied on Day 3 1 Turn the Corning MicroDEN unit OFF disconnect the power cord and remove from the incubator 2 Lightly spray the MicroDEN system with 70 ethanol and place inside the biosafety cabine...

Страница 7: ...e new storage bottle containing differentiation medium Disconnect the empty bottle MPCs and connect the new storage bottle MPCs to the inlet side tubing 8 Transfer the Corning MicroDEN system from the...

Страница 8: ...he flasks back onto the thermoform tray with the corner notch of each flask in the upper right position 10 Repeat Steps 1 9 for a second cold buffer wash NOTE Flasks can be harvested individually by u...

Страница 9: ...anol to disinfect before placing in the biosafety cabinet After use the unit can be cleaned with 70 ethanol or 10 bleach solution The unit should be stored upright in a cool dry environment when not i...

Страница 10: ...riginal material or workmanship This warranty does not cover motor brushes fuses light bulbs batteries or damage to paint or finish Claims for transit damage should be filed with the transportation ca...

Отзывы: