background image

13

12

10. Place the flasks back onto the tray with the corner notch of each flask in the 

upper right position once medium has completely drained into the harvest 
bottle.

NOTE

: Corning® MicroDEN® flasks can be harvested individually by unclamping 

one of the tubes from the pump to stop flow. Liquid will flow through the 
clamped tube.

Cold Buffer Washes

1.  Add 26 mL of cold buffer (4°C) to the left side inlet storage bottle.

To unscrew the cap, turn the bottle while keeping the cap and tubing set 
in place so the tubes don’t twist (Figure 1). Take care not to touch the dip 
tubes to any non-sterile surface.

2.  Insert tubing set and place the cap back onto the bottle. While pressing  

down on the tubing set, rotate the orange cap to tighten the assembly to 
the bottle (Figure 1). Place the bottle back into the thermoform tray.

3.  Select “Harvest” using the function button on the Corning® MicroDEN®

 

unit 

to perfuse cold buffer into the flasks.

4.  Remove the flasks from the thermoform tray and slide the flasks side-to-

side as buffer fills the flasks (Figure 6). 

5.  Allow perfusion to continue when the inlet bottle is empty of buffer. Air will 

perfuse into the flasks.

6.  When air starts to perfuse, rotate the flasks so that the outlet port is down 

and the inlet port is up (Figure 5). The corner notch of each flask should be 
at the lower left position.

Reversing flask orientation allows the liquid to drain into the harvest bottle.

7.  Continue to slide the flasks side-to-side as liquid drains from the flasks into 

the harvest bottle and air is perfused into the flasks.

8.  Stop perfusion by pressing the MicroDEN

 

unit function button when buffer is 

drained from the flasks.

9.  Place the flasks back onto the thermoform tray with the corner notch of each 

flask in the upper right position.

10. Repeat Steps 1-9 for a second cold buffer wash.
NOTE:

 Flasks can be harvested individually by unclamping one of the tubes from 

the pump to stop flow. Liquid will flow through the clamped tube.

Non-adherent and most adherent iDCs are transferred from the MicroDEN flasks 
to the harvest bottle after two cold buffer washes. Additional washes can be 
incorporated if the remaining strongly adherent cells are desired. Additionally, 
the flasks, consumable kit, or the entire unit with consumable kit can be placed 
on a platform shaker to aid detaching strongly adherent cells. Alternatively, the 
flasks or consumable kit can be gently vortexed. Do not place the MicroDEN

 

unit 

on the vortex.

The dendritic cell product is collected in the harvest bottle after washing 
with cold buffer. The harvest bottle can then be removed from the rest of the 
consumable by disconnecting the MPCs. 

Helpful Harvest Tips

When beginning to perfuse liquid, or when the tubing is filled with air, angle 
the left side inlet MPCs upwards so that air is completely removed from the 
MPCs (Figure 4). If this is not performed, air will be trapped in the MPCs and 
can cause an air bubble to perfuse into the flask.

When beginning to perfuse air, angle the left side inlet MPCs downwards so 
that liquid is drained from the MPCs (Figure 7).

Disposing of Consumables

The entire consumable kit (thermoform tray, flasks, tubing, connectors, and bottles) 
are designated as single-use and should be appropriately discarded after use.

NOTE:

 Tubing 

connections 
not shown

Corner notch on 
flask in lower 
left position

Outlet ports 
are down

Inlet ports 
are up

Slide the flasks 
side to side

Figure 6. 

Slide the flasks side-to-side as medium drains into the harvest bottle.

Figure 7. 

MPCs angled downward to drain liquid.

Содержание MicroDEN

Страница 1: ...hnical information visit www corning com lifesciences or call 800 492 1110 Outside the United States call 1 978 442 2200 or contact your local Corning sales office Corning Incorporated Life Sciences 8...

Страница 2: ...autoclave this unit Do not perform the sterilization cycle of an incubator with the MicroDEN unit inside Do not use acetone to clean the unit Always use the provided external power supply to power the...

Страница 3: ...nts 1 Corning MicroDEN unit Cat No 431276 requires standard electrical outlet 2 Corning consumable set Cat No 431277 3 Corning storage bottle Cat No 431278 4 Enriched monocytes MOs 5 Differentiation m...

Страница 4: ...t a wall to prevent disturbing the cultured cells The cell culture flasks may be seeded with either enriched MOs or with peripheral blood mononuclear cells PBMCs where the MOs are enriched through pla...

Страница 5: ...ow through the unclamped tube Method 2 Pump Perfusion 1 Unclamp the pumps by pressing upwards at the dotted triangle on the pump head 2 Place the tubing onto the pump rollers The colored tubing marker...

Страница 6: ...emptied on Day 3 1 Turn the Corning MicroDEN unit OFF disconnect the power cord and remove from the incubator 2 Lightly spray the MicroDEN system with 70 ethanol and place inside the biosafety cabine...

Страница 7: ...e new storage bottle containing differentiation medium Disconnect the empty bottle MPCs and connect the new storage bottle MPCs to the inlet side tubing 8 Transfer the Corning MicroDEN system from the...

Страница 8: ...he flasks back onto the thermoform tray with the corner notch of each flask in the upper right position 10 Repeat Steps 1 9 for a second cold buffer wash NOTE Flasks can be harvested individually by u...

Страница 9: ...anol to disinfect before placing in the biosafety cabinet After use the unit can be cleaned with 70 ethanol or 10 bleach solution The unit should be stored upright in a cool dry environment when not i...

Страница 10: ...riginal material or workmanship This warranty does not cover motor brushes fuses light bulbs batteries or damage to paint or finish Claims for transit damage should be filed with the transportation ca...

Отзывы: