Pipet

MPCs for bottle 
change over

Power cord 
connection

Method 1: Positive Pressure

1.  Unclamp the pumps by pressing upwards at the dotted triangle on the  

pump head.

2.  Place the tubing onto the pump rollers. The colored tubing marker should 

be to the left of each pump and match the colored flag next to each pump 
(Figure 2).

3.  Leave the pumps unclamped.
4.  Connect a pipettor to the sterile filter of the inlet bottle and use the “pipette 

down” button to pressurize the inlet bottle (Figure 3). Cell solution will flow 
from the seeding bottle, through the tubing, and into the flasks.

5.  Tilt the seeding bottle to allow all the liquid to flow into the dip tubes. This 

ensures that all of the cell solution is perfused into the flasks.

6.  Stop pipetting when the seeding bottle is empty. The inlet side tubing will 

still be filled with cell solution.

7.  If cell solution enters the outlet tubing, “pipet up” to reverse flow.
8.  Clamp the tubing onto the pumps by pressing down on the pump head clamp.
9.  The tubing holder on each side of the display may be used to help secure the 

tubing in place.

NOTE: 

Corning® MicroDEN® flasks can be filled separately by clamping one of 

the tubes in the pump to stop flow. Liquid will flow through the unclamped tube.

Method 2: Pump Perfusion

1.  Unclamp the pumps by pressing upwards at the dotted triangle on the pump 

head.

2.  Place the tubing onto the pump rollers. The colored tubing marker should 

be to the left of each pump and match the colored flag next to each pump 
(Figure 2).

3.  Clamp the tubing onto the pumps by pressing down on the pump head 

clamp.

4.  Turn the Corning® MicroDEN® unit “ON,” and select “Seed” using the function 

button on the unit.

5.  The pumps will perfuse the cell solution from the seeding bottle into the flasks.
6.  Tilt the seeding bottle to allow all the liquid to aspirate into the dip tubes. 

This ensures that all of the cell solution is perfused into the flasks.

7.  Stop perfusion by pressing the MicroDEN

 

unit function button when the 

seeding bottle is empty. The inlet side tubing will still be filled with cell 
solution.

NOTE:

 Corning MicroDEN flasks can be filled separately by unclamping one of 

the tubes to stop flow. Liquid will flow through the clamped tube.

Helpful Seeding Tips

During seeding, tilt the inlet bottle to allow all the liquid to flow into the dip 
tubes. This ensures that all of the cell solution is perfused into the flasks.

Tubing 
holder

Thermoform 
tray

Inlet ports

Outlet ports

Color 
markers 
match

ON/OFF 
switch

Function 
button

Pumps shown 
in clamped 
position

Figure 2. 

Set up for cell seeding.

Figure 3. 

Pressurizing the inlet bottle.

Pipet up 
or down

Содержание MicroDEN

Страница 1: ...hnical information visit www corning com lifesciences or call 800 492 1110 Outside the United States call 1 978 442 2200 or contact your local Corning sales office Corning Incorporated Life Sciences 8...

Страница 2: ...autoclave this unit Do not perform the sterilization cycle of an incubator with the MicroDEN unit inside Do not use acetone to clean the unit Always use the provided external power supply to power the...

Страница 3: ...nts 1 Corning MicroDEN unit Cat No 431276 requires standard electrical outlet 2 Corning consumable set Cat No 431277 3 Corning storage bottle Cat No 431278 4 Enriched monocytes MOs 5 Differentiation m...

Страница 4: ...t a wall to prevent disturbing the cultured cells The cell culture flasks may be seeded with either enriched MOs or with peripheral blood mononuclear cells PBMCs where the MOs are enriched through pla...

Страница 5: ...ow through the unclamped tube Method 2 Pump Perfusion 1 Unclamp the pumps by pressing upwards at the dotted triangle on the pump head 2 Place the tubing onto the pump rollers The colored tubing marker...

Страница 6: ...emptied on Day 3 1 Turn the Corning MicroDEN unit OFF disconnect the power cord and remove from the incubator 2 Lightly spray the MicroDEN system with 70 ethanol and place inside the biosafety cabine...

Страница 7: ...e new storage bottle containing differentiation medium Disconnect the empty bottle MPCs and connect the new storage bottle MPCs to the inlet side tubing 8 Transfer the Corning MicroDEN system from the...

Страница 8: ...he flasks back onto the thermoform tray with the corner notch of each flask in the upper right position 10 Repeat Steps 1 9 for a second cold buffer wash NOTE Flasks can be harvested individually by u...

Страница 9: ...anol to disinfect before placing in the biosafety cabinet After use the unit can be cleaned with 70 ethanol or 10 bleach solution The unit should be stored upright in a cool dry environment when not i...

Страница 10: ...riginal material or workmanship This warranty does not cover motor brushes fuses light bulbs batteries or damage to paint or finish Claims for transit damage should be filed with the transportation ca...

Отзывы: