2
3
3.0 Installation
Upon unpacking the Corning® LSE™ Digital Dry Bath, inspect for damages .
Shipping damage is the responsibility of the carrier . Inspect that the
following are present: instruction manual, power cord, block lifter .
Select a location that is dry and not subject to drafts or moving air from
heating or air conditioning vents, or air blown by other equipment . Place
the unit on a flat, non-flammable surface . Allow sufficient room around the
unit for access and cooling . Six inches minimum on all sides is suggested .
Plug the unit into a properly grounded outlet . By using the block lifter,
insert the sample block into the well . The unit is now ready for use .
4.0 Controls
On/Off
rocker switch, behind the unit: turns primary power on and off
Start/Stop
LED, red: starts or stops the unit, illuminates when unit is in
heating mode and off in temperature set mode
Set Temp
LED, red: illuminates when unit is in temperature set mode and
off when in heating mode
Heating
LED, red: illuminates when unit is in heating mode and unit is
actually applying heat to the block, on continuously during heat-up and
cycles on and off when the unit is at the set temperature
Timer
LED, green: sets or selects timer mode, illuminates while running in
timer mode
Up
arrow key: raises set temperature or time when unit is in set mode
Down
arrow key: lowers set temperature or time when unit is in set mode
Mode
key: switches the unit between heating mode, set temperature mode,
or timer mode
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
1.0 Specifications
Temperature range
A5°C to 150°C
Temperature display resolution
0 .1°C 4 digit LED
Temperature uniformity
± 0 .2°C (at 37°C in block)
Temperature accuracy
± 0 .3°C
Temperature controller
Microprocessor – user calibratable
Timer
1 to 999 minutes in 1 minute
increments or continuous
Operating altitude
Not to exceed 2,000 meters
Control
PI microprocessor controller
I/O
RS232 unidirectional
Dimensions (W x D x H)
7 .9 x 10 .4 x 3 .3 in . (20 x 26 .5 x 8 .3 cm)
Weight
4 .8 lbs ./2 .2 kg
Electrical requirements
D1100
115V
50/60Hz, 0 .97A, fuse 1 .6 AT
D1100
230V
50/60Hz, 1 .48A, fuse 0 .63 AT
D1200
115V
50/60Hz, 1 .45A, fuse 2 .0 AT
D1200
230V
50/60Hz, 0 .72A, fuse 1 .0 AT
2.0 Safety Precautions
Do not
use this product in an explosive environment .
Do not
use in the presence of flammable or combustible material .
Do not
heat substances that react violently when heated .
Do not
touch block when hot or when unit is heating
(use block lifter) .
Do not
touch area around block or block well when unit is hot .
Do not
spill liquids into the well area or into the unit side
vent holes .
Connect unit
only to a properly grounded outlet .
Caution: HOT Surfaces that can burn or cause injury.
Use of this product in any manner not specified by the manufacturer or
modification of the product may cause injury and/or may void the warranty .