3.2.1
Warnings, Precautions and Notes
1. The use of disinfecting solutions for an exterior instrument wipe will
not sterilize the equipment and is not recommended.
2. Do not “Peel Pack” handpieces for sterilization. Sterilization in a sealed pouch
traps moisture which can cause damage.
3. The suction control valve must be in the fully open position during sterilization.
4. Do not use handpieces while warm. Allow adequate time for cooling prior to
use. Cool by exposure to room temperature. Operation of a handpiece that is not
completely cool or dry may decrease performance and/or reliability.
Handpieces may be processed in a pre-vacuum steam sterilizer (Steam Pre-
vacuum). Individually wrap handpiece and handpiece cord. Follow the
recommended minimum sterilization exposure times listed below.
*PRECAUTION: An eight (8) minute minimum dry cycle must be run
on all handpieces and attachments every time the product is sterilized.
Failure to use a dry cycle on the products may lead to reduced product
performance or premature product failure.
Method
Cycle
Minimum
Temperature
Minimum
Exposure
Minimum*
Dry Cycle
Steam (wrapped)
Pre-Vacuum
270°F (132°C)
4 minutes
8 minutes