1

1.0

INTRODUCTION

These shaver handpieces come in two styles; no-button and 2-button functionality.
They work with a variety of ConMed Linvatec shaver blades and burs and all are used
in conjunction with several drive consoles; the Advantage® System (D3000 Series),
E9000® System (E9000 Series), and the PowerPro® System (PRO2000 Series) (all
with software version 7.0 or higher).

These shaver handpieces are reusable and are supplied non-sterile and must
be sterilized prior to the first use and after every use.

1.1

Intended Use

The shaver handpieces are designed to operate ConMed Linvatec shaver blades
and burs for resection of bone and soft tissue in arthroscopic, foot, hand,
orthopedic and otolaryngological surgical procedures.

1.2

Operating Principle

The shaver handpiece is electrically powered by the surgical console to provide high
speed

,

 rotary force to the accessory (shaver blade, shaver bur or attachment) for tissue

resection. The shaver handpiece is controlled by operating buttons on the handpiece
or using a footswitch. A suction source is connected to the shaver handpiece to draw
tissue into the shaver blade and remove tissue from the surgical site.

1.3

Environmental Directives

WEEE Directive [2002/96/EC] on Waste Electrical and Electronic
Equipment. This statement only applies to European countries with regard to
the Waste Electrical and Electric Equipment (WEEE) European Directive.

The WEEE symbol on the product or its packaging indicates that this
product must not be disposed of with other waste. Instead, it is your responsibility to
dispose of your waste equipment by handing it over to a designated collection point
for the recycling of Waste Electrical and Electronic Equipment. The separate
collection and recycling of your waste equipment at the time of disposal will help
conserve natural resources and ensure that it is recycled in a manner that protects
human health and the environment. For more information about where you can drop
off your medical equipment at the end of its useful life for recycling, please contact
ConMed Linvatec.

W41-261-004AB.book  Page 1  Thursday, September 18, 2014  9:28 AM

Содержание Linvatec Advantage Turbo D9920

Страница 1: ...ConMed Linvatec Shaver Handpieces D9920 D9924 W41 261 004AB book Page 1 Thursday September 18 2014 9 28 AM...

Страница 2: ...ated transmitted transcribed stored in a retrieval system or translated into any human or computer language in any form or by any means electronic mechanical magnetic manual or otherwise or disclosed...

Страница 3: ...ion Symbols 7 1 6 Handpiece Descriptions 8 1 7 Preoperative Functional Test 9 2 0 CLEANING AND STERILIZING 2 1 Cleaning and Washer Sanitizer Warnings and Precautions 10 3 0 Handpiece Cleaning Instruct...

Страница 4: ...Table of Contents Page ii 4 3 Handpiece Environmental Requirements 15 5 0 CUSTOMER SERVICE 5 1 Assistance 16 W41 261 004AB book Page ii Thursday September 18 2014 9 28 AM...

Страница 5: ...led by operating buttons on the handpiece or using a footswitch A suction source is connected to the shaver handpiece to draw tissue into the shaver blade and remove tissue from the surgical site 1 3...

Страница 6: ...accessory Damage to the equipment is possible if these instructions are not followed NOTE A note is added to provide additional focused information This equipment is designed for use by medical profes...

Страница 7: ...y clean and re sterilize this single use device has not been established and subsequent re use may adversely affect the performance safety and or sterility of the device 8 After use shaver blades burs...

Страница 8: ...ment for proper operation Ensure all attachments and accessories are correctly and completely attached to the handpiece Perform the required Performance Tests for each associated handpiece prior to ea...

Страница 9: ...mance is not increased with force Excessive force or using shaver blades or burs as a lever can cause damage to the device including permanent deformation shedding of metal wear motor seizure and over...

Страница 10: ...to Speed Increase button see step 4 Handpiece Descriptions on page 8 for more information Handpiece ON OFF button Will also convert to Speed Decrease button see step 4 Handpiece Descriptions on page...

Страница 11: ...Law restricts this device to sale by or on the order of a physician Manufacturer Consult instructions for use Non sterile Date of Manufacture No user service recommended Refer servicing to qualified C...

Страница 12: ...he blade to ensure it is locked To remove the blade rotate and hold the collet in the UNLOCK direction and remove the blade 2 Suction Port Valve Manually controls suction flow Move towards the front o...

Страница 13: ...ed and held for approximately 1 second the handpiece will run in the preset forward mode for as long as the ON OFF button is held down Release the ON OFF button and the handpiece will resume running i...

Страница 14: ...eased handpiece performance 6 For aluminum surfaces a neutral pH agent should be used To prevent corrosion avoid contact with strong alkaline solutions pH over 10 5 or agents containing iodine or chlo...

Страница 15: ...binding the internal mechanism of the collet Clean until all debris is removed 5 Manipulate all moving parts of the handpiece to ensure all debris is removed If not clean again until all debris is rem...

Страница 16: ...that is not completely cool or dry may decrease performance and or reliability Handpieces may be processed in a pre vacuum steam sterilizer Steam Pre vacuum Individually wrap handpiece and handpiece...

Страница 17: ...he system will accept interference received including interference that might cause undesirable operation If interference occurs separate the instruments by moving one away from the other For more inf...

Страница 18: ...ndpieces Operating Speed 500 12 000 rpm Forward Reverse 500 4 000 cpm Single and Multi turn Oscillate Torque Nominal 35 in oz 24 7 N cm Length 6 1 in 15 5 cm Diameter Approx 1 25 in tear drop 3 2 cm W...

Страница 19: ...al Requirements Environmental Conditions Operating Storage and Transport Temperature Relative Humidity Atmospheric Pressure 75 30 Non Condensing 93 10 Non Condensing W41 261 004AB book Page 15 Thursda...

Страница 20: ...splayed on the box and included on all paperwork Refer to this number if making inquiries about the repair status Please call ConMed Linvatec Customer Service and provide the following information to...

Страница 21: ...W41 261 004AB book Page 1 Thursday September 18 2014 9 28 AM...

Страница 22: ...525 French Road Utica NY 13502 5994 USA 727 392 6464 Customer Service 1 800 237 0169 FAX 727 399 5256 International FAX 1 727 397 4540 email customer_service conmed com www conmed com W41 261 004AB b...

Отзывы: