9

Speed button. This enables the surgeon to change the handpiece operating
speed from the sterile field. If no button is pressed for 1.5 seconds, the
console beeps twice and returns to normal button functionality.

5. ON/OFF Button - Press and release to turn the shaver handpiece on and

off.

Controlling window indexing using the buttons on the D9924: The
following operation applies only when the handpiece is NOT running. When
the handpiece ON/OFF button is pressed and held, the console beeps once
and the blade window starts to rotate (index) slowly at 6 rpm. When the
window position of the blade is found (generally window completely open or
window completely shut), release the ON/OFF button. The handpiece will
return the blade window to this location when it is stopped.

6. Turbo Mode Using Ultra Series Shaver Blades and D4000/D4000A Drive 

Console with console software revision V 2.0 or higher: The following applies 
to the D9924, Shaver handpieces and only when it is actively running in 
oscillate mode. When the ON/OFF button is pressed and held for approximately 
1 second, the handpiece will run in the preset forward mode for as long as the 
ON/OFF button is held down. Release the ON/OFF button and the handpiece 
will resume running in the oscillate mode. When the DIRECTION SELECT 
button is pressed and held for approximately 1 second, the handpiece will run in 
the preset reverse mode for as long as the DIRECTION SELECT button is held 
down. Release the DIRECTION SELECT button and the handpiece will 
resume running in the oscillate mode.

1.7

Preoperative Functional Test

Prior to each use, perform the following preoperative functional test:

1. Insert a bur or blade into the handpiece by aligning one of the slots in the collet

with the key on the blade/bur hub. Gently pull on the cutting accessory to
ensure it is properly seated.

2. Verify that the console properly recognizes the handpiece and cutting accessory. 

3. Press the direction select button or footswitch to ensure that it functions

properly. 

4. Run the handpiece for less than 5 seconds at a speed less than 1500 rpm to

observe any abnormal noises, vibrations or heat rise.

5. If any operating difficulties occur, return the handpiece for service.

W41-261-004AB.book  Page 9  Thursday, September 18, 2014  9:28 AM

Содержание Linvatec Advantage Turbo D9920

Страница 1: ...ConMed Linvatec Shaver Handpieces D9920 D9924 W41 261 004AB book Page 1 Thursday September 18 2014 9 28 AM...

Страница 2: ...ated transmitted transcribed stored in a retrieval system or translated into any human or computer language in any form or by any means electronic mechanical magnetic manual or otherwise or disclosed...

Страница 3: ...ion Symbols 7 1 6 Handpiece Descriptions 8 1 7 Preoperative Functional Test 9 2 0 CLEANING AND STERILIZING 2 1 Cleaning and Washer Sanitizer Warnings and Precautions 10 3 0 Handpiece Cleaning Instruct...

Страница 4: ...Table of Contents Page ii 4 3 Handpiece Environmental Requirements 15 5 0 CUSTOMER SERVICE 5 1 Assistance 16 W41 261 004AB book Page ii Thursday September 18 2014 9 28 AM...

Страница 5: ...led by operating buttons on the handpiece or using a footswitch A suction source is connected to the shaver handpiece to draw tissue into the shaver blade and remove tissue from the surgical site 1 3...

Страница 6: ...accessory Damage to the equipment is possible if these instructions are not followed NOTE A note is added to provide additional focused information This equipment is designed for use by medical profes...

Страница 7: ...y clean and re sterilize this single use device has not been established and subsequent re use may adversely affect the performance safety and or sterility of the device 8 After use shaver blades burs...

Страница 8: ...ment for proper operation Ensure all attachments and accessories are correctly and completely attached to the handpiece Perform the required Performance Tests for each associated handpiece prior to ea...

Страница 9: ...mance is not increased with force Excessive force or using shaver blades or burs as a lever can cause damage to the device including permanent deformation shedding of metal wear motor seizure and over...

Страница 10: ...to Speed Increase button see step 4 Handpiece Descriptions on page 8 for more information Handpiece ON OFF button Will also convert to Speed Decrease button see step 4 Handpiece Descriptions on page...

Страница 11: ...Law restricts this device to sale by or on the order of a physician Manufacturer Consult instructions for use Non sterile Date of Manufacture No user service recommended Refer servicing to qualified C...

Страница 12: ...he blade to ensure it is locked To remove the blade rotate and hold the collet in the UNLOCK direction and remove the blade 2 Suction Port Valve Manually controls suction flow Move towards the front o...

Страница 13: ...ed and held for approximately 1 second the handpiece will run in the preset forward mode for as long as the ON OFF button is held down Release the ON OFF button and the handpiece will resume running i...

Страница 14: ...eased handpiece performance 6 For aluminum surfaces a neutral pH agent should be used To prevent corrosion avoid contact with strong alkaline solutions pH over 10 5 or agents containing iodine or chlo...

Страница 15: ...binding the internal mechanism of the collet Clean until all debris is removed 5 Manipulate all moving parts of the handpiece to ensure all debris is removed If not clean again until all debris is rem...

Страница 16: ...that is not completely cool or dry may decrease performance and or reliability Handpieces may be processed in a pre vacuum steam sterilizer Steam Pre vacuum Individually wrap handpiece and handpiece...

Страница 17: ...he system will accept interference received including interference that might cause undesirable operation If interference occurs separate the instruments by moving one away from the other For more inf...

Страница 18: ...ndpieces Operating Speed 500 12 000 rpm Forward Reverse 500 4 000 cpm Single and Multi turn Oscillate Torque Nominal 35 in oz 24 7 N cm Length 6 1 in 15 5 cm Diameter Approx 1 25 in tear drop 3 2 cm W...

Страница 19: ...al Requirements Environmental Conditions Operating Storage and Transport Temperature Relative Humidity Atmospheric Pressure 75 30 Non Condensing 93 10 Non Condensing W41 261 004AB book Page 15 Thursda...

Страница 20: ...splayed on the box and included on all paperwork Refer to this number if making inquiries about the repair status Please call ConMed Linvatec Customer Service and provide the following information to...

Страница 21: ...W41 261 004AB book Page 1 Thursday September 18 2014 9 28 AM...

Страница 22: ...525 French Road Utica NY 13502 5994 USA 727 392 6464 Customer Service 1 800 237 0169 FAX 727 399 5256 International FAX 1 727 397 4540 email customer_service conmed com www conmed com W41 261 004AB b...

Отзывы: