904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 5
In addition, please compare the accessories you receive with the standard accessories list below. If
an item is missing, please notify CONMED Customer Service.
Standard Accessories:
• Operator’s Manual
• Power Cord
• Pneumatic Footswitch
• Automatic Activation Device
Please contact CONMED Customer Service to purchase the following accessories:
• Replacement Filters
• Tubing Accessories
• Fluid Trap
1.3 Operational Information
The operational information contained in this section is intended for the customer review of
regulatory issues. The information pertains to the use of the products both domestically and
internationally:
1. CONMED ClearView™ Smoke Evacuation System(s) complies with IEC60601.1
electrical specifications in the following systems:
100/120 VAC
50/60 Hz, 220/240 VAC 50/60 Hz
2. Type of protection against electrical shock (UL 60601-1, Clause 5.1): Class I
3. Degree of protection against electric shock (UL 60601-1, Clause 5.2); Type CF Applied Part
4. Degree of protection against ingress of water (UL 60601-1, Clause 5.3): IPX1
5. Method of sterilization or disinfection recommended by CONMED (UL 60601-1,
Clause 5.4):
Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution
or soapy water. Wipe dry with a clean cloth. Do not steam sterilize.
6. Degree of safety of application in the presence of flammable anesthetic mixture with air or
with oxygen or nitrous oxide (UL 60601-1, Clause 5.5): Not Suitable
7. Mode of operation (UL 60601-1, Clause 5.6): Continuous
8. Upon request, CONMED will provide the following:
Service and Repair Instructions, including Circuit Diagrams and Parts List
9. The fuses on the circuit board are to be serviced by an authorized CONMED techni-
cian as follows:
100/120 VAC, 50/60 Hz use 10 Amp 250 Volt Fuse (Slo-Blo)
220/240 VAC, 50/60 Hz use 8 Amp 250 Volt Fuse (Slo-blo)