904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual 

Page 12

•  FILTER LIFE INDICATOR 

    (Membrane Control Panel)

The filter life indicator on the membrane control panel provides a visual indication of the status of 

the life of the filter in use.  The filter life indicator for the ClearView™ Smoke Evacuation System 

will automatically adjust according to the flow setting selected. 

 

 

Low/Lap Flow Setting = 35 hours of filter life

 

 

Medium Flow Setting = 24 hours of filter life

 

 

High/Turbo Flow Setting = 18 hours of filter life 

The ClearView™ Filter Life Indicator is factory set.  All filter life timing is automatic.

Reading the Filter Life Indicator:

Install an unused ClearView™ Filter into the system per the installation instructions contained in 

this operator’s manual.  When the system is turned on, the filter life indicator with light up the 

leftmost GREEN LED, indicating 100% filter life. The indicator will progress thru subsequent 

GREEN LEDs, to an AMBER LED as time elapses and begin flashing RED to indicate that the 

filter has expended its useful life and requires replacement.

When the maximum filter life is consumed and the smoke evacuator is not powered off for 

greater than six (6) hours or the main power is disconnected – a new filter is required to 

activate the smoke evacuator and make operational.

•  FUSES

    (circuit board)

Two 10 AMP fuses (8 AMP for 220/240 ClearView™ Smoke Evacuation Systems) are located on 

the circuit board within the housing of the system.  It electrically protects both the system and the 

operator from damage or injury.  If the system is overheated or if there is an electrical surge in the 

electrical system, fuses will break and the system will not operate.  

When the Service light illuminates, please contact CONMED Customer Service for system 

service instructions.

Содержание ClearView 60-8585-120

Страница 1: ...ClearView Smoke Evacuation System Operator s Manual...

Страница 2: ...A The revision level of this manual is specified by the highest revision letter found on either the inside front cover or enclosed errata pages if any Manual Number 904568 REVA Unit Serial Number ____...

Страница 3: ...ions and Warnings 2 0 OPERATING INSTRUCTIONS 10 2 1 System Controls 2 2 Filter Instructions 2 3 Set up and Operation 2 4 Specifications 3 0 MAINTENANCE 21 3 1 General Maintenance Information 3 2 Clean...

Страница 4: ...ion Air filter whose high tech patented U S Patent 5874052 design captures particulates and micro organisms from 1 to 2 microns at an efficiency of 99 999 The third stage filtration uses the highest g...

Страница 5: ...VAC 50 60 Hz 220 240 VAC 50 60 Hz 2 Type of protection against electrical shock UL 60601 1 Clause 5 1 Class I 3 Degree of protection against electric shock UL 60601 1 Clause 5 2 Type CF Applied Part...

Страница 6: ...cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the inte...

Страница 7: ...ult The CONMED filters and single use accessories are completely disposable Please dispose of according to your local codes or regulations and hospital policy These filters may be disposed of or incin...

Страница 8: ...uid If fluid is expected to be aspirated to the ClearView Filter fluid collection devices must be installed with the vacuum hose assembly Failure to install a fluid collection device could cause filte...

Страница 9: ...nvironmental ambient temperature 14 F to140 F 10 C to 60 C Storage environmental relative humidity 10 to 75 There are no user serviceable components in the ClearView Smoke Evacuation System s Refer se...

Страница 10: ...QUALIFIED SERVICE PERSONNEL DANGER CAUTION RISK OF EXPLOSION IF USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS WARNING CAUTION TYPE CF APPLIED PART IPX1 PROTECTION AGAINST INGRESS OF WATER AS DETAILED...

Страница 11: ...g the suction ON membrane switch to illuminate yellow STANDBY LED Turn the system main power off O by unplugging the power cord from the appliance inlet on the unit or the receptacle in the wall SUCTI...

Страница 12: ...ystem is turned on the filter life indicator with light up the leftmost GREEN LED indicating 100 filter life The indicator will progress thru subsequent GREEN LEDs to an AMBER LED as time elapses and...

Страница 13: ...ice is not intended for evacuation of fluid If fluid is expected to be aspirated using the Filter or the CONMED ClearView Smoke Evacuation System fluid collection devices must be installed with the va...

Страница 14: ...2015 All rights reserved 904568 Rev A CONMED ClearView Smoke Evacuation System Operator s Manual Page 14 Figure 1 Unit Control Panel...

Страница 15: ...mensions H x W x D inches 6 x 11 x 15 5 Dimensions H x W x D centimeters 15 2 x 27 9 x 39 4 Weight lbs 4 3 Noise Level dBA nominal 55 0 dBA Footswitch Pneumatic Standard Automatic Control Activation Y...

Страница 16: ...it is used in such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 The ClearView uses RF energy only for its internal function Therefore it...

Страница 17: ...hat of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical...

Страница 18: ...meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vici...

Страница 19: ...equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 3 5 d P v1 80 kHz to 800 MHz 3 5 d P v1 800 kHz to 2 5 GHz 3 5 d P...

Страница 20: ...oper mating cleanliness and absence of damage to the filter inlet Obvious external or internal damage to the system 3 4 Troubleshooting the System see below Maintenance Section 3 0 PROBLEM POTENTIAL C...

Страница 21: ...ClearView Smoke Evacuation System use it to properly return your unit If you do not have the original packing material ask Customer Service for advice on how to pack the unit for return shipment Step...

Страница 22: ...NMED by purchaser for the specific system giving rise to the liability Purchaser agrees to indemnify and hold CONMED harmless from and against all liabilities claims and demands of third parties of an...

Страница 23: ...matters involving the validity interpretation and application of this agreement shall be controlled by the laws of the State of New York Using any filter not offered by CONMED may cause damage to the...

Страница 24: ...e Evacuation System Operator s Manual Page 24 Distributed by ConMed Corporation 525 French Road Utica NY 13502 USA 1 800 448 6506 U S Only or 1 315 797 8375 International www conmed com Manufactured b...

Отзывы: