904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual 

Page 20

3.1 

General Maintenance Information

This section contains information for ordinary upkeep of the CONMED ClearView™ Smoke 

Evacuation System.  While the system has been designed and manufactured to high industry 

standards, it is recommended that periodic inspection and performance testing be performed by 

a qualified Biomedical Technician to ensure continued safe and effective operation.

3.2 Cleaning

Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution or soapy water. 

Wipe dry with a clean cloth. Do not steam sterilize.

 

3.3 

Periodic Inspection

The ClearView™ Smoke Evacuation System should be visually inspected at least every year.  This 

inspection should include checks for:

•  Damage to the power cord.

•  Damage to the power plug or power inlet module. 

•  Proper mating, cleanliness and absence of damage to the filter inlet.

•  Obvious external or internal damage to the system.

3.4 

Troubleshooting the System – see below.

Maintenance

Section 3.0

PROBLEM                            POTENTIAL CAUSE       CORRECTIVE ACTION

1. Smoke Evacuation System

is ON but suction is minimal

or none.

2. Smoke Evacuation System

does not function even

though suction ON button is

depressed.

1. Filter is not seated

completely.

2. Filter is clogged.

3. Vacuum hose or tube is

clogged.

4. Motor/blower is obstructed.

1. Not plugged into an

electrical outlet.

2. Fuses are blown.

3. Electronic system failure.

4. Filter life has expired or

invalid filter inserted.

1. Re-install Filter, press firmly into place and fully 

engage clip.

2. Replace filter with a genuine CONMED Filter.

3. Replace vacuum hose or tube with genuine

CONMED products.

4. Call BioMed or CONMED Customer

Service at 1.800.448.6506 or +1.315.797.8375.

1. Check power outlet and connection to rear or

side panel of the machine.

2./3.  Call BioMed or CONMED Customer

Service at 1.800.448.6506 or +1.315.797.8375.

4. Replace filter with a genuine CONMED Filter.

Содержание ClearView 60-8585-120

Страница 1: ...ClearView Smoke Evacuation System Operator s Manual...

Страница 2: ...A The revision level of this manual is specified by the highest revision letter found on either the inside front cover or enclosed errata pages if any Manual Number 904568 REVA Unit Serial Number ____...

Страница 3: ...ions and Warnings 2 0 OPERATING INSTRUCTIONS 10 2 1 System Controls 2 2 Filter Instructions 2 3 Set up and Operation 2 4 Specifications 3 0 MAINTENANCE 21 3 1 General Maintenance Information 3 2 Clean...

Страница 4: ...ion Air filter whose high tech patented U S Patent 5874052 design captures particulates and micro organisms from 1 to 2 microns at an efficiency of 99 999 The third stage filtration uses the highest g...

Страница 5: ...VAC 50 60 Hz 220 240 VAC 50 60 Hz 2 Type of protection against electrical shock UL 60601 1 Clause 5 1 Class I 3 Degree of protection against electric shock UL 60601 1 Clause 5 2 Type CF Applied Part...

Страница 6: ...cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the inte...

Страница 7: ...ult The CONMED filters and single use accessories are completely disposable Please dispose of according to your local codes or regulations and hospital policy These filters may be disposed of or incin...

Страница 8: ...uid If fluid is expected to be aspirated to the ClearView Filter fluid collection devices must be installed with the vacuum hose assembly Failure to install a fluid collection device could cause filte...

Страница 9: ...nvironmental ambient temperature 14 F to140 F 10 C to 60 C Storage environmental relative humidity 10 to 75 There are no user serviceable components in the ClearView Smoke Evacuation System s Refer se...

Страница 10: ...QUALIFIED SERVICE PERSONNEL DANGER CAUTION RISK OF EXPLOSION IF USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS WARNING CAUTION TYPE CF APPLIED PART IPX1 PROTECTION AGAINST INGRESS OF WATER AS DETAILED...

Страница 11: ...g the suction ON membrane switch to illuminate yellow STANDBY LED Turn the system main power off O by unplugging the power cord from the appliance inlet on the unit or the receptacle in the wall SUCTI...

Страница 12: ...ystem is turned on the filter life indicator with light up the leftmost GREEN LED indicating 100 filter life The indicator will progress thru subsequent GREEN LEDs to an AMBER LED as time elapses and...

Страница 13: ...ice is not intended for evacuation of fluid If fluid is expected to be aspirated using the Filter or the CONMED ClearView Smoke Evacuation System fluid collection devices must be installed with the va...

Страница 14: ...2015 All rights reserved 904568 Rev A CONMED ClearView Smoke Evacuation System Operator s Manual Page 14 Figure 1 Unit Control Panel...

Страница 15: ...mensions H x W x D inches 6 x 11 x 15 5 Dimensions H x W x D centimeters 15 2 x 27 9 x 39 4 Weight lbs 4 3 Noise Level dBA nominal 55 0 dBA Footswitch Pneumatic Standard Automatic Control Activation Y...

Страница 16: ...it is used in such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 The ClearView uses RF energy only for its internal function Therefore it...

Страница 17: ...hat of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical...

Страница 18: ...meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vici...

Страница 19: ...equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 3 5 d P v1 80 kHz to 800 MHz 3 5 d P v1 800 kHz to 2 5 GHz 3 5 d P...

Страница 20: ...oper mating cleanliness and absence of damage to the filter inlet Obvious external or internal damage to the system 3 4 Troubleshooting the System see below Maintenance Section 3 0 PROBLEM POTENTIAL C...

Страница 21: ...ClearView Smoke Evacuation System use it to properly return your unit If you do not have the original packing material ask Customer Service for advice on how to pack the unit for return shipment Step...

Страница 22: ...NMED by purchaser for the specific system giving rise to the liability Purchaser agrees to indemnify and hold CONMED harmless from and against all liabilities claims and demands of third parties of an...

Страница 23: ...matters involving the validity interpretation and application of this agreement shall be controlled by the laws of the State of New York Using any filter not offered by CONMED may cause damage to the...

Страница 24: ...e Evacuation System Operator s Manual Page 24 Distributed by ConMed Corporation 525 French Road Utica NY 13502 USA 1 800 448 6506 U S Only or 1 315 797 8375 International www conmed com Manufactured b...

Отзывы: