904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 20
3.1
General Maintenance Information
This section contains information for ordinary upkeep of the CONMED ClearView™ Smoke
Evacuation System. While the system has been designed and manufactured to high industry
standards, it is recommended that periodic inspection and performance testing be performed by
a qualified Biomedical Technician to ensure continued safe and effective operation.
3.2 Cleaning
Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution or soapy water.
Wipe dry with a clean cloth. Do not steam sterilize.
3.3
Periodic Inspection
The ClearView™ Smoke Evacuation System should be visually inspected at least every year. This
inspection should include checks for:
• Damage to the power cord.
• Damage to the power plug or power inlet module.
• Proper mating, cleanliness and absence of damage to the filter inlet.
• Obvious external or internal damage to the system.
3.4
Troubleshooting the System – see below.
Maintenance
Section 3.0
PROBLEM POTENTIAL CAUSE CORRECTIVE ACTION
1. Smoke Evacuation System
is ON but suction is minimal
or none.
2. Smoke Evacuation System
does not function even
though suction ON button is
depressed.
1. Filter is not seated
completely.
2. Filter is clogged.
3. Vacuum hose or tube is
clogged.
4. Motor/blower is obstructed.
1. Not plugged into an
electrical outlet.
2. Fuses are blown.
3. Electronic system failure.
4. Filter life has expired or
invalid filter inserted.
1. Re-install Filter, press firmly into place and fully
engage clip.
2. Replace filter with a genuine CONMED Filter.
3. Replace vacuum hose or tube with genuine
CONMED products.
4. Call BioMed or CONMED Customer
Service at 1.800.448.6506 or +1.315.797.8375.
1. Check power outlet and connection to rear or
side panel of the machine.
2./3. Call BioMed or CONMED Customer
Service at 1.800.448.6506 or +1.315.797.8375.
4. Replace filter with a genuine CONMED Filter.