904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual
Page 19
[ ]
[ ]
[ ]
Table 4
Recommended Separation Distance Between
Portable and Mobile RF Communications Equipment and the Model @ 3 Vrms
The model ClearView™ is intended for use in the electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the model ClearView™ can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the model ClearView™ as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
3.5
d
=
√P
v
1
80 kHz to 800 MHz
3.5
d
=
√P
v
1
800 kHz to 2.5 GHz
3.5
d
=
√P
v
1
0.01
0.12
0.12
0.23
0.1
0.34
0.34
0.74
1
1.7
1.7
2.3
10
3.7
3.7
7.4
100
11.7
11.7
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance (D)
in meters (M) can be estimated using the equation applicable to the frequency of the transmitter, where (P) is the
maximum output rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structure, objects, and people.