34
3.
On fully automatic Medical IntelliSwitch® systems, switch the Regulator Status toggle switch down – on the side to be
replaced - to indicate that the regulator is “Out of Service”.
4.
Close the inlet and outlet isolation valves for the delivery regulator to be replaced, see figure 26.
Figure 26
5.
Loosen the inlet and outlet tube fittings and remove the ‘faulty’ regulator, see figure 27.
Figure 27
Содержание IntelliSwitch 57 Series